FDA Adverse Event Injury Summary report: N

GLIDEWELL HT IMPLANT Ø3.5 X 11.5 MM

MDR report key: 23296513 · Received October 15, 2025

Report

Report Number
3011649314-2025-01273
Event Type
Injury
Date Received
October 15, 2025
Date of Event
May 15, 2025
Report Date
January 2, 2026
Manufacturer
PRISMATIK DENTALCRAFT, INC.
Product Code
DZE
PMA / PMN Number
K143353
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
DENTIST
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

CORRECTED INFORMATION: D4. MANUFACTURER REFERENCE: (B)(4).

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION: D9. THE DEVICE HAS BEEN RECEIVED AND THE INVESTIGATION HAS BEEN COMPLETED. THE RESULTS ARE AS FOLLOWS: DHR RESULTS: THE DHR WAS REVIEWED FOR GLIDEWELL HT IMPLANT LOT# 6242130 AND THERE WAS NO EVIDENCE DISCOVERED TO INDICATE THAT A PRODUCT DEFECT OR NON-CONFORMITY CONTRIBUTED TO THE ISSUE. THE PART MET ALL THE CRITERIA CALLED FOR IN THE PRODUCTION ROUTER. STOCK PRODUCT REVIEWED RESULTS: A REVIEW OF STOCK PRODUCT WAS PERFORMED FOR GLIDEWELL HT IMPLANT LOT# 6242130 AND FOUND NO ADDITIONAL PRODUCT IN STOCK. INVESTIGATION METHODS/RESULTS: THE DEVICE WAS RETURNED BUT NOT IN THE ORIGINAL PACKAGE. THE IMPLANT WAS VERIFIED TO BE A GLIDEWELL HT IMPLANT Ø3.5 X 11.5 MM (70-1189-IMP0006) USING THE RADIOGRAPHIC TEMPLATE (MKT-013579 REV 1 PK-4500230-112023). THERE WAS NO DEFECT OR NON-CONFORMITY OBSERVED AND THE THREADS WERE INTACT. MATTER WAS OBSERVED IN THE TREADING OF THE IMPLANT. THE COMPLAINT IS VERIFIED BASED ON THE RETURNED PART BUT CANNOT CONFIRM THE FAILURE MODE. THERE WAS NO EVIDENCE FOUND THAT INDICATED THAT THE REPORTED ISSUE WAS CAUSED BY THE DEVICE ITSELF. ROOT CAUSE DESCRIPTION: "LACK OF PRIMARY STABILITY" AND "FAILURE TO OSSEOINTEGRATE" ARE COMMON COMPLAINTS IN REGARD TO IMPLANT FAILURE. THIS OCCURS WHEN THE PATIENT'S BONE DOES NOT INTEGRATE WITH THE IMPLANT SURFACE. THE POSSIBLE RESPONSES TO THIS COMPLAINT COULD BE ATTRIBUTED TO VARIOUS CAUSES. ALTHOUGH THE ROOT CAUSE FOR LACK OF PRIMARY STABILITY AND FAILURE TO OSSEOINTEGRATE ARE INCONCLUSIVE AND SPECIFIC TO EACH CASE, PROBABLE CAUSES COULD BE DUE TO INSUFFICIENT BONE OR POOR BONE QUALITY; EITHER THE BONE WAS TOO SOFT, OR THE OPERATOR ERRED IN CREATING AN OSTEOTOMY BIGGER THAN THE SIZE OF THE IMPLANT DIAMETER. PREMATURE LOADING, PATIENT'S HEALTH, PERI-IMPLANTITIS, SMOKING, AND LACK OF ORAL HYGIENE MAY ALSO BE CONTRIBUTING FACTORS. PER THE REPORTED INFORMATION, THE PATIENT HAS TYPE III BONE QUALITY AND FAIR ORAL HYGIENE. IFU-012631 REV 1 (GLIDEWELL HT IMPLANT SYSTEM IFU) CONTAINS THE FOLLOWING STATEMENT IN THE CONTRAINDICATIONS SECTION: "GLIDEWELL HT IMPLANTS SHOULD NOT BE PLACED IN PATIENTS DISCOVERED TO BE MEDICALLY UNFIT FOR THE INTENDED TREATMENT. PRIOR TO CLINICAL INTERVENTION, PROSPECTIVE PATIENTS MUST BE THOROUGHLY EVALUATED FOR ALL KNOWN RISK FACTORS AND CONDITIONS RELATED TO ORAL SURGICAL PROCEDURES AND SUBSEQUENT HEALING. CONTRAINDICATIONS INCLUDE BUT ARE NOT LIMITED TO: VASCULAR CONDITIONS, UNCONTROLLED DIABETES, CLOTTING DISORDERS, ANTICOAGULANT THERAPY, METABOLIC BONE DISEASE, CHEMOTHERAPY OR RADIATION THERAPY, CHRONIC PERIODONTAL INFLAMMATION, INSUFFICIENT SOFT TISSUE COVERAGE, METABOLIC OR SYSTEMIC DISORDERS ASSOCIATED WITH WOUND AND/OR BONE HEALING, USE OF PHARMACEUTICALS THAT INHIBIT OR ALTER NATURAL BONE REMODELING, ANY DISORDERS WHICH INHIBIT A PATIENT'S ABILITY TO MAINTAIN ADEQUATE DAILY ORAL HYGIENE, UNCONTROLLED PARAFUNCTIONAL HABITS, INSUFFICIENT HEIGHT AND/OR WIDTH OF BONE, AND INSUFFICIENT INTERARCH SPACE." IFU-012631 REV 1 (GLIDEWELL HT IMPLANT SYSTEM IFU) CONTAINS THE FOLLOWING STATEMENT IN THE SURGICAL PROCEDURES UNDER THE PRECAUTION SECTION: "MINIMIZING TISSUE DAMAGE IS CRUCIAL TO SUCCESSFUL IMPLANT OSSEOINTEGRATION. IN PARTICULAR, CARE SHOULD BE TAKEN TO ELIMINATE SOURCES OF INFECTION, CONTAMINANTS, SURGICAL AND THERMAL TRAUMA. RISK OF OSSEOINTEGRATION FAILURE INCREASES AS TISSUE TRAUMA INCREASES. FOR BEST RESULTS, PLEASE OBSERVE THE FOLLOWING PRECAUTIONS: ALL DRILLING PROCEDURES SHOULD BE PERFORMED AT 2000 RPM OR LESS UNDER CONTINUAL, COPIOUS IRRIGATION. ALL SURGICAL INSTRUMENTS USED MUST BE IN GOOD CONDITION AND SHOULD BE USED CAREFULLY TO AVOID DAMAGE TO IMPLANTS OR OTHER COMPONENTS. IMPLANTS SHOULD BE PLACED WITH SUFFICIENT STABILITY; HOWEVER, EXCESSIVE INSERTION TORQUE MAY RESULT IN IMPLANT FRACTURE, OR FRACTURE OR NECROSIS OF THE IMPLANT SITE. THE PROPER SURGICAL PROTOCOL SHOULD BE STRICTLY ADHERED TO." IFU-012631 REV 1 (GLIDEWELL HT IMPLANT SYSTEM IFU) CONTAINS THE FOLLOWING STATEMENT IN THE WARNINGS SECTION: "THE IMPLANT SITE SHOULD BE INSPECTED FOR ADEQUATE BONE BY RADIOGRAPHS, PALPATIONS AND VISUAL EXAMINATION. DETERMINE THE LOCATION OF NERVES AND OTHER VITAL STRUCTURES AND THEIR PROXIMITY TO THE IMPLANT SITE BEFORE ANY DRILLING TO AVOID POTENTIAL INJURY, SUCH AS PERMANENT NUMBNESS TO THE LOWER LIP AND CHIN." IFU-012631 REV 1 (GLIDEWELL HT IMPLANT SYSTEM IFU) CONTAINS THE FOLLOWING STATEMENT IN THE WARNINGS SECTION: "ABSOLUTE SUCCESS CANNOT BE GUARANTEED. FACTORS SUCH AS INFECTION, DISEASE, AND INADEQUATE BONE QUALITY AND/OR QUANTITY CAN RESULT IN OSSEOINTEGRATION FAILURES FOLLOWING SURGERY OR INITIAL OSSEOINTEGRATION." MANUFACTURER REFERENCE: (B)(4).

Additional Manufacturer Narrative · 0

THE REPORTED DEVICE HAS NOT YET BEEN RETURNED. AN INVESTIGATION WILL BE CARRIED OUT AND A SUPPLEMENTAL REPORT WILL BE SUBMITTED UPON COMPLETION OF THE INVESTIGATION. MANUFACTURER REFERENCE: (B)(4).

Description of Event or Problem · 0

IT WAS REPORTED BY A HEALTHCARE PROFESSIONAL THAT THE GLIDEWELL HT IMPLANT FAILED. THE PATIENT'S BONE QUALITY WAS REPORTED AS TYPE III AND THEIR ORAL HYGIENE AS FAIR. THE PATIENT PRESENTED FOR IMPLANT PLACEMENT ON TOOTH POSITION #8 ON (B)(6) 2025. THE PATIENT RETURNED FOR A FOLLOW-UP VISIT ON (B)(6) 2025 AND UPON EXAMINATION, THE PROVIDER NOTED THERE WAS BONE LOSS AROUND THE IMPLANT. THE PROVIDER DETERMINED THAT THE REPORTED DEVICE LACKED PRIMARY STABILITY/FAILED TO OSSEOINTEGRATE. THE DEVICE WAS EXPLANTED THAT DAY AND REPLACED WITH A PRODUCT SIMILAR TO THE COMPLAINT PRODUCT. THE SYMPTOMS RESOLVED AFTER IMPLANT REMOVAL AND THERE WAS NO PERMANENT PATIENT INJURY. BONE GRAFTING WAS PERFORMED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2288154 GLIDEWELL HT IMPLANT Ø3.5 X 11.5 MM GLIDEWELL HT DZE PRISMATIK DENTALCRAFT, INC. 70-1189-IMP0006 6242130

Patients

Seq Age Sex Outcome Treatment
1 66 YR Male Required Intervention