FDA Adverse Event Malfunction Summary report: N

HEATER-COOLER SYSTEM

MDR report key: 23288158 · Received October 15, 2025

Report

Report Number
9611109-2025-90431
Event Type
Malfunction
Date Received
October 15, 2025
Date of Event
September 15, 2025
Report Date
October 14, 2025
Manufacturer
LIVANOVA DEUTSCHLAND GMBH
Product Code
DWC
UDI-DI
04033817901099
Removal / Correction Number
Z-2076/2081-2015
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

A.1.-A.5. THERE IS NO REPORT OF ANY PATIENT INVOLVEMENT. G.5. THE HEATER-COOLER 16-02-80 IS NOT DISTRIBUTED IN THE USA, AND IT IS SIMILAR TO HEATER-COOLER 16-02-85, WHICH IS DISTRIBUTED IN THE USA (510(K) NUMBER: K220635). H.9. LIVANOVA DEUTSCHLAND IMPLEMENTED A FIELD SAFETY NOTICE FOR DISINFECTION AND CLEANING OF HEATER-COOLER DEVICES. THE Z NUMBER IS Z-2076/2081-2015. H11: LIVANOVA DEUTSCHLAND MANUFACTURES THE 3T HEATER COOLER THE USER REQUIRED DEEP DISINFECTION PROCEDURE. IF ANY ADDITIONAL INFORMATION PERTINENT TO THE REPORTED EVENT IS RECEIVED, IT WILL BE PROVIDED IN A SUPPLEMENTAL REPORT.

Description of Event or Problem · 0

LIVANOVA DEUTSCHLAND RECEIVED A REPORT THAT A 3T HEATER COOLER RESULTED POSITIVE TO MYCOBACTERIUM CHIMAERA. THERE IS NO REPORT OF ANY PATIENT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2596177 HEATER-COOLER SYSTEM CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS DWC LIVANOVA DEUTSCHLAND GMBH HEATER COOLER 3T 04033817901099

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown