HEATER-COOLER SYSTEM
MDR report key: 23288158
·
Received October 15, 2025
Report
- Report Number
- 9611109-2025-90431
- Event Type
- Malfunction
- Date Received
- October 15, 2025
- Date of Event
- September 15, 2025
- Report Date
- October 14, 2025
- Manufacturer
- LIVANOVA DEUTSCHLAND GMBH
- Product Code
- DWC
- UDI-DI
- 04033817901099
- Removal / Correction Number
- Z-2076/2081-2015
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 0
A.1.-A.5. THERE IS NO REPORT OF ANY PATIENT INVOLVEMENT. G.5. THE HEATER-COOLER 16-02-80 IS NOT DISTRIBUTED IN THE USA, AND IT IS SIMILAR TO HEATER-COOLER 16-02-85, WHICH IS DISTRIBUTED IN THE USA (510(K) NUMBER: K220635). H.9. LIVANOVA DEUTSCHLAND IMPLEMENTED A FIELD SAFETY NOTICE FOR DISINFECTION AND CLEANING OF HEATER-COOLER DEVICES. THE Z NUMBER IS Z-2076/2081-2015. H11: LIVANOVA DEUTSCHLAND MANUFACTURES THE 3T HEATER COOLER THE USER REQUIRED DEEP DISINFECTION PROCEDURE. IF ANY ADDITIONAL INFORMATION PERTINENT TO THE REPORTED EVENT IS RECEIVED, IT WILL BE PROVIDED IN A SUPPLEMENTAL REPORT.
Description of Event or Problem · 0
LIVANOVA DEUTSCHLAND RECEIVED A REPORT THAT A 3T HEATER COOLER RESULTED POSITIVE TO MYCOBACTERIUM CHIMAERA. THERE IS NO REPORT OF ANY PATIENT INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2596177 | HEATER-COOLER SYSTEM | CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS | DWC | LIVANOVA DEUTSCHLAND GMBH | HEATER COOLER 3T | 04033817901099 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |