FDA Adverse Event
Malfunction
Summary report: N
HEATER-COOLER SYSTEM
MDR report key: 23287689
·
Received October 15, 2025
Report
- Report Number
- 9611109-2025-90429
- Event Type
- Malfunction
- Date Received
- October 15, 2025
- Date of Event
- September 15, 2025
- Report Date
- May 20, 2026
- Manufacturer
- LIVANOVA DEUTSCHLAND GMBH
- Product Code
- DWC
- UDI-DI
- 04033817901099
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
Additional Manufacturer Narrative · 0
A.1.-A.5. THERE IS NO REPORT OF ANY PATIENT INVOLVEMENT. G.5. THE HEATER-COOLER 16-02-80 IS NOT DISTRIBUTED IN THE USA, AND IT IS SIMILAR TO HEATER-COOLER 16-02-85, WHICH IS DISTRIBUTED IN THE USA (510(K) NUMBER: K220635). H11: LIVANOVA DEUTSCHLAND MANUFACTURES THE 3T HEATER COOLER BASED ON THE AVAILABLE INFORMATION, COOLING COIL EROSION, ALLOWING WATER INTO COMPRESSOR, IS SUSPECTED, IF ANY ADDITIONAL INFORMATION PERTINENT TO THE REPORTED EVENT IS RECEIVED, IT WILL BE PROVIDED IN A SUPPLEMENTAL REPORT.
Description of Event or Problem · 0
LIVANOVA DEUTSCHLAND RECEIVED A REPORT THAT A 3T HEATER COOLER TRIPS MAINS CIRCUIT BREAKERS IN THEATRES 20 SECONDS AFTER POWER UP. THERE WAS NO PATIENT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 609277 | HEATER-COOLER SYSTEM | CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS | DWC | LIVANOVA DEUTSCHLAND GMBH | HEATER COOLER 3T | 04033817901099 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |