FDA Adverse Event Malfunction Summary report: N

HF SENSOR DELIVERY SYSTEM

MDR report key: 23285902 · Received October 15, 2025

Report

Report Number
3004936110-2025-06206
Event Type
Malfunction
Date Received
October 15, 2025
Date of Event
September 8, 2025
Report Date
November 3, 2025
Manufacturer
ABBOTT MEDICAL
Product Code
MOM
UDI-DI
05414734509237
PMA / PMN Number
P100045
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

NO DEVICE ANALYSIS RESULTS ARE AVAILABLE BECAUSE THE DEVICE REMAINS IMPLANTED. A REVIEW OF THE DHR WAS PERFORMED AND RULED OUT THE POSSIBILITY OF A MANUFACTURING RELATED ISSUE CAUSING OR CONTRIBUTING TO THE REPORTED COMPLAINT ISSUE. PER THE IFU, RIGHT HEART CATHETERIZATION IS REQUIRED FOR SYSTEM BASELINE (MEAN PRESSURE) CALIBRATION AND MAY BE NEEDED TO RECALIBRATE THE BASELINE (MEAN PRESSURE) IN ORDER TO CONTINUE TO USE THE SYSTEM.

Additional Manufacturer Narrative · 0

THE INVESTIGATION CODES ALONG WITH INVESTIGATION RESULTS WILL BE PROVIDED IN A SUBSEQUENT SUBMISSION.

Description of Event or Problem · 0

ADDITIONAL INFORMATION WAS PROVIDED ON OCTOBER 7, 2025, CONFIRMING THAT THE RIGHT HEART CATHETERIZATION PROCEDURE WAS NOT PERFORMED PRIMARILY TO CHECK THE CARDIOMEMS PRESSURES BUT WAS BEING PERFORMED AS A ROUTINE CARE FOR THIS PATIENT.

Description of Event or Problem · 0

A RIGHT HEART CATHETERIZATION WAS PERFORMED TO CONFIRM THE VALIDITY OF THE PRESSURE SENSOR READINGS. THE SENSOR WAS RECALIBRATED AND THE MEAN WAS DECREASED BY 5 MMHG. READINGS WERE OBSERVED UNDER THE MANUFACTURER'S PATIENT CARE NETWORK DATABASE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2733423 HF SENSOR DELIVERY SYSTEM SYSTEM, HEMODYNAMIC, IMPLANTABLE MOM ABBOTT MEDICAL CM2000 7963314 05414734509237

Patients

Seq Age Sex Outcome Treatment
1 71 YR Male Required Intervention