ANIMAS INSULIN CARTRIDGE
Report
- Report Number
- 2531779-2011-08359
- Event Type
- Malfunction
- Date Received
- November 8, 2011
- Report Date
- October 17, 2011
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K032257
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- OTHER
Narratives
CARTRIDGE LOT # B201671. THE CARTRIDGE HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. ANIMAS SHALL CONDUCT A LOT PROFILE TEST ON THE CARTRIDGE LOT AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
(B)(4). DEVICE EVALUATION: NO CARTRIDGE WAS RETURNED; A RETAINED SAMPLE WAS EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2012 WITH THE FOLLOWING FINDINGS: THE CARTRIDGE PASSED VISUAL INSPECTION, NO DAMAGE OR DEFECTS WERE NOTED TO THE LUER CONNECTION OR O-RINGS. A LEAK TEST WAS PERFORMED WITH NO FAILURES BEING OBSERVED; NO LEAKS WERE OBSERVED FROM THE LUER CONNECTION, O-RINGS, OR ANYWHERE ELSE IN THE CARTRIDGE.
THE PATIENT REPORTED THAT HER BLOOD GLUCOSE LEVEL WAS 400 MG/DL ON OCTOBER 15. THE READING DOES NOT FIT ANIMAS CRITERIA OF A SERIOUS DIABETIC INJURY. WHEN SHE REMOVED THE CARTRIDGE AND TUBING TO CHANGE THEM SHE PUSHED ON THE PLUNGER MANUALLY AND STATES INSULIN LEAKED OUT OF THE BODY OF THE CARTRIDGE AND INTO HER HAND. SHE IS UNSURE WHETHER THE PLUNGER WAS COCKED AT THE TIME OF PUSHING. SHE DENIES ANY LEAKAGE FOUND IN THE CARTRIDGE COMPARTMENT. SHE DISCARDED THE CARTRIDGE BUT IS CONCERNED AS TO WHY IT WOULD LEAK WHEN SHE PUSHES THE PLUNGER. THE ANIMAS REPRESENTATIVE INSTRUCTED THE PATIENT TO CYCLE THE CARTRIDGE ONLY TWICE VERY SLOWLY. THE PATIENT REPORTEDLY CYCLES THE CARTRIDGE FOUR TIMES BEFORE FILLING IT WITH INSULIN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN CARTRIDGE | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ANM IR1200/1250/2020/OTP CART |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 65 YR |