FDA Adverse Event Malfunction Summary report: N

FLOWTRON EXCEL TRIO PUMP

MDR report key: 2328160 · Received October 21, 2011

Report

Report Number
1000381138-2011-00030
Event Type
Malfunction
Date Received
October 21, 2011
Date of Event
September 1, 2011
Report Date
September 22, 2011
Manufacturer
ARJOHUNTLEIGH
Product Code
JOW
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS BEING FILED UNDER (B)(4) BY ARJOHUNTLEIGH ON BEHALF OF THE MANUFACTURER ARJOHUNTLEIGH, A BRANCH OF ARJO (B)(4). THE EVAL OF THE DEVICE TO DATE HAS BEEN CARRIED OUT BY A REP OF THE MANUFACTURER'S SALES AND SERVICE UNIT SUBSIDIARY DIVISIONS, NOT A DIRECT EMPLOYEE OF THE MANUFACTURER. THE INFO CONTAINED IN THIS INITIAL REPORT IS BASED UPON THE INFO PROVIDED TO THE MANUFACTURER. ADDITIONAL INFO WILL BE PROVIDED FOLLOWING THE CONCLUSION OF THE MANUFACTURER'S INVESTIGATION.

Description of Event or Problem · 1

CONTACT STATES THAT THE FLOWTRON TRIO WAS IN USE DURING A SURGICAL PROCEDURE WITH PLACEMENT ON THE FLOOR UNDERNEATH THE SURGICAL TABLE. AN OPERATING ROOM PERSONNEL (SCRUB TECH) ACCIDENTALLY HIT THE ON/OFF SWITCH WITH THEIR FOOT WHICH CAUSED THE SWITCH TO POP OFF CAUSING SPARKS FROM THE PUMP AS THE SURGICAL PROCEDURE WAS IN PROGRESS. THE TRIO SYSTEM WAS TAKEN OUT OF SERVICE IMMEDIATELY AND REPLACED WITH ANOTHER TRIO SYSTEM DURING COMPLETION OF THE PROCEDURE. NO PT OR STAFF INJURIES WERE INCURRED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FLOWTRON EXCEL TRIO PUMP JOW ARJOHUNTLEIGH 512003

Patients

Seq Age Sex Outcome Treatment
1 Other