FDA Adverse Event Injury Summary report: N

INSET I

MDR report key: 23276022 · Received October 14, 2025

Report

Report Number
3003442380-2025-14924
Event Type
Injury
Date Received
October 14, 2025
Date of Event
July 17, 2025
Report Date
September 30, 2025
Manufacturer
UNOMEDICAL DEVICES S.A. DE C.V.
Product Code
FPA
UDI-DI
05705244022386
PMA / PMN Number
K032854
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

INITIAL AND FINAL MDR (B)(4) DEVICE 1 OF 6. ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY. THE REFERENCE SAMPLES FOR THE LOT 6011396 HAVE ALREADY BEEN PREVIOUSLY TESTED IN THE COMPLAINT (B)(4) ON 25-JUL- 2025. TEST RESULTS: VISUAL TEST ACCORDING TO WORK INSTRUCTION (WI) VERSION 3.0 ON REFERENCE SAMPLES, 10 SAMPLES OUT 10 SAMPLES PASSED THE TEST. FUNCTIONAL FLOW TEST 1 ACCORDING TO WI VERSION 2.0 ON REFERENCE SAMPLES, 10 SAMPLES OUT 10 SAMPLES PASSED THE TEST. DEVICE HISTORY RECORD (DHR) REVIEW: THE LOT 6011396 WAS MANUFACTURED ACCORDING TO THE WI VERSION 120 IN THE LINE 138, ON 29/JAN/2025, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DHR SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WERE IDENTIFIED RELATED TO THE MALFUNCTION REPORTED, NO MAINTENANCE EVENTS WERE RECORDED. THRESHOLD ANALYSIS: A QUERY WAS RUN ON 30-SEP-2025 AGAINST "FINAL REPORTING DECISION EQUAL "SERIOUS INJURY" AND "DEATH", "LOT NUMBER" CRITERIA EQUAL "6011396". THE COUNT OF COMPLAINT IS 2 WHICH IS BELOW 3. NO FURTHER STATISTICAL TRENDING ANALYSIS IS REQUIRED. CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: AS A RESULT OF THE FOLLOWING: NO PHYSICAL SAMPLES WERE RETURNED. NO NON-CONFORMANCE (NC) RAISED DURING PRODUCTION RELATED TO COMPLAINT CODE, TWO COMPLAINTS THRESHOLDS IDENTIFIED FOR THE LOT IN QUESTION AND SERIOUS INJURY/DEATH, NO FURTHER ACTIONS ARE REQUIRED. AS A RESULT OF THE POST MARKET SURVEILLANCE ACTIVITIES, THIS COMPLAINT HAS BEEN EVALUATED AS A TYPE 4 (ESCALATED TO CAPA): THIS IS A COMPLAINT WHICH IS BEING ADDRESSED AS PART OF A CORRECTIVE AND PREVENTIVE ACTION (CAPA) 1768101 AUTOSOFT 90, KINKED SOFT CANNULA ISSUES WHICH HAS BEEN OPENED AS RESULT OF TRENDING ACTIVITIES. THIS IS A COMPLAINT WHICH IS BEING ADDRESSED AS PART OF A CAPA 1768101: "AUTOSOFT 90, KINKED SOFT CANNULA ISSUES WHICH HAS BEEN OPENED AS A RESULT OF TRENDING ACTIVITIES. THIS COMPLAINT IS A REPORTABLE WITH VALID LOT NUMBER/PRODUCT CODE. CAPA DETERMINATION RESULTS: THIS COMPLAINT FALLS UNDER THE SCOPE OF THE CAPA 1768101: "AUTOSOFT 90, KINKED SOFT CANNULA ISSUES" AND WILL COVER INSET I (AUTOSOFT XC DHF-13.8 & 13.13) AND INSET II (AUTOSOFT 90 DHF-14 AND 14.3) PRODUCTS. ROOT CAUSE OF PROBLEM: THE ROOT CAUSE HAS BEEN IDENTIFIED AS: METHOD, MANPOWER, MEASUREMENT, MACHINE: CORRECTIVE ACTION AS A RESULT OF THE INVESTIGATION: THE ACTION PLAN AND DEADLINES IS AS FOLLOWED: THE FOLLOWING ACTIONS WERE ALREADY IMPLEMENTED IN THE NC 1515780 · 1. INCLUDE THE DEFECT IN DOCUMENT (B) (WORK INSTRUCTION INSET LINE) · 2. IMPROVE GUARDS OF THE CONVEYORS · 3. UPDATE TRAYS FOR THE STOCK OF CANNULAS · 4. UPDATE DOCUMENT (B)(4) (QUALITY SPECIFICATION FOR CATHETER FIXTURE FOR SKEWED CATHETERS) FOR INCLUDE THE HANDLING OF THE CATHETER DURING THE PROCESS. · 5. UPDATE THE DOCUMENT (B)(4) (WORK INSTRUCTION INSET LINE) TO INCLUDE THE PREVENTIVE ACTIONS IMPLEMENTED IN THE PROCESS (TRAYS) A REMAINING ACTION (TO ADDRESS DESIGN ROOT CAUSE) AND DEADLINES IS AS FOLLOWED: · ADD CYLINDERS TO THE LID TO MAINTAIN IN POSITION THE NEEDLE AND CANNULA DURING TRANSPORTATION AND PRIOR USE (TW (B)(4) 30/SEP/2025).

Additional Manufacturer Narrative · 0

SUPPLEMENTAL REPORT 01 - (B)(4) - MDR 3003442380-2025-14924. ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY. THE REFERENCE SAMPLES FOR THE LOT 6011396 HAVE ALREADY BEEN PREVIOUSLY TESTED IN THE COMPLAINT (B)(4) ON 25-JUL- 2025. TEST RESULTS: VISUAL TEST ACCORDING TO WI VERSION 3.0 ON REFERENCE SAMPLES, 10 SAMPLES OUT 10 SAMPLES PASSED THE TEST. FUNCTIONAL FLOW TEST 1 ACCORDING TO (B)(4) VERSION 2.0 ON REFERENCE SAMPLES, 10 SAMPLES OUT 10 SAMPLES PASSED THE TEST. DEVICE HISTORY RECORD (DHR) REVIEW: THE LOT 6011396 WAS MANUFACTURED ACCORDING TO THE WI VERSION 120 IN THE LINE 138, ON 29/JAN/2025, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DHR SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WERE IDENTIFIED RELATED TO THE MALFUNCTION REPORTED, NO MAINTENANCE EVENTS WERE RECORDED. THRESHOLD ANALYSIS: A QUERY WAS RUN ON 30-SEP-2025 AGAINST "FINAL REPORTING DECISION EQUAL "SERIOUS INJURY" AND "DEATH", "LOT NUMBER" CRITERIA EQUAL "6011396". THE COUNT OF COMPLAINT IS 2 WHICH IS BELOW 3. NO FURTHER STATISTICAL TRENDING ANALYSIS IS REQUIRED. CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: AS A RESULT OF THE FOLLOWING: NO PHYSICAL SAMPLES WERE RETURNED. NO NON-CONFORMANCE (NC) RAISED DURING PRODUCTION RELATED TO COMPLAINT CODE, TWO COMPLAINTS THRESHOLDS IDENTIFIED FOR THE LOT IN QUESTION AND SERIOUS INJURY/DEATH, NO FURTHER ACTIONS ARE REQUIRED. AS A RESULT OF THE POST MARKET SURVEILLANCE ACTIVITIES, THIS COMPLAINT HAS BEEN EVALUATED AS A TYPE 4 (ESCALATED TO CAPA): THIS IS A COMPLAINT WHICH IS BEING ADDRESSED AS PART OF A CORRECTIVE AND PREVENTIVE ACTION (CAPA) 1768101 AUTOSOFT 90, KINKED SOFT CANNULA ISSUES WHICH HAS BEEN OPENED AS RESULT OF TRENDING ACTIVITIES. THIS IS A COMPLAINT WHICH IS BEING ADDRESSED AS PART OF A CAPA 1768101: "AUTOSOFT 90, KINKED SOFT CANNULA ISSUES WHICH HAS BEEN OPENED AS A RESULT OF TRENDING ACTIVITIES. THIS COMPLAINT IS A REPORTABLE WITH VALID LOT NUMBER/PRODUCT CODE. CAPA DETERMINATION RESULTS: THIS COMPLAINT FALLS UNDER THE SCOPE OF THE CAPA 1768101: "AUTOSOFT 90, KINKED SOFT CANNULA ISSUES" AND WILL COVER INSET I (AUTOSOFT XC DHF-13.8 & 13.13) AND INSET II (AUTOSOFT 90 DHF-14 AND 14.3) PRODUCTS. ROOT CAUSE OF PROBLEM: THE ROOT CAUSE HAS BEEN IDENTIFIED AS: METHOD, MANPOWER, MEASUREMENT, MACHINE: CORRECTIVE ACTION AS A RESULT OF THE INVESTIGATION: THE ACTION PLAN AND DEADLINES IS AS FOLLOWED: THE FOLLOWING ACTIONS WERE ALREADY IMPLEMENTED IN THE NC 1515780 1. INCLUDE THE DEFECT IN DOCUMENT 4805074 (WORK INSTRUCTION INSET LINE). 2. IMPROVE GUARDS OF THE CONVEYORS. 3. UPDATE TRAYS FOR THE STOCK OF CANNULAS. 4. UPDATE DOCUMENT 3A02003 (QUALITY SPECIFICATION FOR CATHETER FIXTURE FOR SKEWED CATHETERS) FOR INCLUDE THE HANDLING OF THE CATHETER DURING THE PROCESS. 5. UPDATE THE DOCUMENT 4805074 (WORK INSTRUCTION INSET LINE) TO INCLUDE THE PREVENTIVE ACTIONS IMPLEMENTED IN THE PROCESS (TRAYS). A REMAINING ACTION (TO ADDRESS DESIGN ROOT CAUSE) AND DEADLINES IS AS FOLLOWED: ADD CYLINDERS TO THE LID TO MAINTAIN IN POSITION THE NEEDLE AND CANNULA DURING TRANSPORTATION AND PRIOR USE (DATABASE 1771074- 30/SEP/2025).

Description of Event or Problem · 0

REFERENCE NUMBER (B)(4) EVENT OCCURRED IN THE UNITED STATES. IT WAS REPORTED THAT PATIENT EXPERIENCED SIX EVENTS OF INFUSION SET BENT CANNULA FROM (B)(6) 2025. THE EVENT OCCURRED THREE OR MORE HOURS AFTER INSERTION. THE INSERTION SITE WAS ABDOMEN. ON (B)(6) 2025, PATIENT WENT TO EMERGENCY ROOM AND EVENTUALLY GOT HOSPITALIZED DUE TO HYPERGLYCEMIA. THE BLOOD GLUCOSE LEVEL WAS 400 MG/DL AT THE TIME OF EVENT AND PATIENT GOT TREATED WITH INTRAVENOUS AND FLUIDS OF SALINE AND INSULIN. THE PATIENT RELEASED FROM THE HOSPITAL ON SAME DAY. NO FURTHER INFORMATION AVAILABLE.

Description of Event or Problem · 0

TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
174804 INSET I UNO INSET I 12/6 GREY TCAP 10PK INT FPA UNOMEDICAL DEVICES S.A. DE C.V. 1006922 6011396 05705244022386

Patients

Seq Age Sex Outcome Treatment
1 33 YR Male Required Intervention