FDA Adverse Event
Malfunction
Summary report: N
HEATER-COOLER SYSTEM 3T
MDR report key: 23268818
·
Received October 10, 2025
Report
- Report Number
- 9611109-2025-90423
- Event Type
- Malfunction
- Date Received
- October 10, 2025
- Date of Event
- September 11, 2025
- Report Date
- October 10, 2025
- Manufacturer
- LIVANOVA DEUTSCHLAND
- Product Code
- DWC
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 0
A.1.-A.5. THERE WAS NO PATIENT INVOLVEMENT. G.5. THE HEATER-COOLER 16-02-80 IS NOT DISTRIBUTED IN THE USA AND IT IS SIMILAR TO HEATER-COOLER 16-02-85, WHICH IS DISTRIBUTED IN THE USA (510(K) NUMBER: K220635). H11: LIVANOVA DEUTSCHLAND MANUFACTURES THE HEATER-COOLER 3T SYSTEM. THE INCIDENT OCCURRED IN FRANCE. LIVANOVA INITIATED AN INVESTIGATION. IF ANY ADDITIONAL INFORMATION PERTINENT TO THE REPORTED EVENT IS RECEIVED, IT WILL BE PROVIDED IN A SUPPLEMENTAL REPORT.
Description of Event or Problem · 0
LIVANOVA RECEIVED REPORT THAT A HEATER COOLER SYSTEM 3T DISPLAYED ERROR CODE E17, PUMP 2 USES TOO MUCH POWER. THE ISSUE OCCURRED DURING PRIMING. THERE WAS NO PATIENT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2280597 | HEATER-COOLER SYSTEM 3T | CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS | DWC | LIVANOVA DEUTSCHLAND | HEATER COOLER 3T |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |