BD SYRINGE 50ML LL TIP 1ML
Report
- Report Number
- 1911916-2025-00676
- Event Type
- Malfunction
- Date Received
- October 9, 2025
- Date of Event
- September 17, 2025
- Report Date
- September 25, 2025
- Manufacturer
- BD MEDICAL (BD WEST) MEDICAL SURGICAL
- Product Code
- FMF
- PMA / PMN Number
- K110771
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MD, US
- Reporter Occupation
- PHARMACIST
- Health Professional
- Yes
Narratives
A BROWN RESIDUE WAS REPORTED INSIDE A SYRINGE. TO SUPPORT THE INVESTIGATION, ONE SAMPLE¿INCLUDING TWO PACKAGING BLISTERS¿AND TWO PHOTOGRAPHS WERE SUBMITTED FOR EVALUATION BY THE QUALITY TEAM. UPON VISUAL INSPECTION, THREE DARK BROWN SPECKS OF EMBEDDED DEGRADED RESIN WERE OBSERVED WITHIN THE SYRINGE BARREL. THE PHOTOGRAPHS PROVIDED CORRESPOND TO THE SAMPLE RECEIVED. NO ADDITIONAL DEFECTS OR ABNORMALITIES WERE NOTED. THE PRESENCE OF EMBEDDED DEGRADED RESIN IS TYPICALLY ASSOCIATED WITH STARTUP CONDITIONS OR INTERMITTENT OCCURRENCES DURING THE INJECTION MOLDING PROCESS. DURING THESE TIMES, DEGRADED RESIN MAY DETACH AND BECOME INCORPORATED INTO MOLDED COMPONENTS. A REVIEW OF THE DEVICE HISTORY RECORD FOR MATERIAL NUMBER 309653, LOT 5171906, WAS CONDUCTED. THE REVIEW DID NOT IDENTIFY ANY QUALITY ISSUES DURING PRODUCTION THAT COULD HAVE CONTRIBUTED TO THE REPORTED CONDITION. NO RELATED QUALITY NOTIFICATIONS WERE FOUND. ALL MANUFACTURING PROCESSES AND FINAL INSPECTIONS WERE PERFORMED IN ACCORDANCE WITH SPECIFICATION REQUIREMENTS. TO DATE, NO SIMILAR EVENTS HAVE BEEN REPORTED FOR THIS LOT. BASED ON THE INVESTIGATION AND ANALYSIS OF THE RETURNED SAMPLE, THE CUSTOMER-REPORTED CONDITION HAS BEEN CONFIRMED. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED MONTHLY. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.
IT WAS REPORTED THAT THE BD SYRINGE 50ML LL TIP 1ML HAD FOREIGN MATTER. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL #309653 LOT #5171906 VERBATIM: RCC RECEIVED A COMPLAINT VIA EMAIL. EMAIL(S) ATTACHED A 50ML LL SYRINGE OPENED BY TECHNICIAN FOUND BROWN RESIDUE INSIDE THE SYRINGE LOT - 5171906,309653.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 310865 | BD SYRINGE 50ML LL TIP 1ML | PISTON SYRINGE | FMF | BD MEDICAL (BD WEST) MEDICAL SURGICAL | 5171906 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |