FDA Adverse Event Malfunction Summary report: N

BD SYRINGE 50ML LL TIP 1ML

MDR report key: 23262740 · Received October 9, 2025

Report

Report Number
1911916-2025-00676
Event Type
Malfunction
Date Received
October 9, 2025
Date of Event
September 17, 2025
Report Date
September 25, 2025
Manufacturer
BD MEDICAL (BD WEST) MEDICAL SURGICAL
Product Code
FMF
PMA / PMN Number
K110771
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
PHARMACIST
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A BROWN RESIDUE WAS REPORTED INSIDE A SYRINGE. TO SUPPORT THE INVESTIGATION, ONE SAMPLE¿INCLUDING TWO PACKAGING BLISTERS¿AND TWO PHOTOGRAPHS WERE SUBMITTED FOR EVALUATION BY THE QUALITY TEAM. UPON VISUAL INSPECTION, THREE DARK BROWN SPECKS OF EMBEDDED DEGRADED RESIN WERE OBSERVED WITHIN THE SYRINGE BARREL. THE PHOTOGRAPHS PROVIDED CORRESPOND TO THE SAMPLE RECEIVED. NO ADDITIONAL DEFECTS OR ABNORMALITIES WERE NOTED. THE PRESENCE OF EMBEDDED DEGRADED RESIN IS TYPICALLY ASSOCIATED WITH STARTUP CONDITIONS OR INTERMITTENT OCCURRENCES DURING THE INJECTION MOLDING PROCESS. DURING THESE TIMES, DEGRADED RESIN MAY DETACH AND BECOME INCORPORATED INTO MOLDED COMPONENTS. A REVIEW OF THE DEVICE HISTORY RECORD FOR MATERIAL NUMBER 309653, LOT 5171906, WAS CONDUCTED. THE REVIEW DID NOT IDENTIFY ANY QUALITY ISSUES DURING PRODUCTION THAT COULD HAVE CONTRIBUTED TO THE REPORTED CONDITION. NO RELATED QUALITY NOTIFICATIONS WERE FOUND. ALL MANUFACTURING PROCESSES AND FINAL INSPECTIONS WERE PERFORMED IN ACCORDANCE WITH SPECIFICATION REQUIREMENTS. TO DATE, NO SIMILAR EVENTS HAVE BEEN REPORTED FOR THIS LOT. BASED ON THE INVESTIGATION AND ANALYSIS OF THE RETURNED SAMPLE, THE CUSTOMER-REPORTED CONDITION HAS BEEN CONFIRMED. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED MONTHLY. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD SYRINGE 50ML LL TIP 1ML HAD FOREIGN MATTER. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL #309653 LOT #5171906   VERBATIM: RCC RECEIVED A COMPLAINT VIA EMAIL. EMAIL(S) ATTACHED A 50ML LL SYRINGE OPENED BY TECHNICIAN FOUND BROWN RESIDUE INSIDE THE SYRINGE LOT - 5171906,309653.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
310865 BD SYRINGE 50ML LL TIP 1ML PISTON SYRINGE FMF BD MEDICAL (BD WEST) MEDICAL SURGICAL 5171906

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown