FDA Adverse Event Malfunction Summary report: N

TRILEAP LOWER EXTREMITY ANATOMIC PLATING SYSTEM

MDR report key: 23260096 · Received October 9, 2025

Report

Report Number
MW5177169
Event Type
Malfunction
Date Received
October 9, 2025
Date of Event
September 26, 2025
Report Date
October 7, 2025
Manufacturer
CROSSROADS EXTREMITY SYSTEMS LLC
Product Code
HWC
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
MI, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Description of Event or Problem · 0

LOCKING SCREW FAILED TO ENGAGE IN AT LEAST ONE OF THE DISTAL HOLES OF THE PLATE DURING ATTEMPTED FIXATION. THE LOCKING MECHANISM DID NOT SEAT OR THREAD PROPERLY INTO THE PLATE, CONSISTENT WITH PRIOR REPORTED ISSUES. THE SCREW WAS EXCHANGED AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT PATIENT HARM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
303576 TRILEAP LOWER EXTREMITY ANATOMIC PLATING SYSTEM SCREW, FIXATION, BONE HWC CROSSROADS EXTREMITY SYSTEMS LLC

Patients

Seq Age Sex Outcome Treatment
1 65 YR Female