FDA Adverse Event
Malfunction
Summary report: N
HEATER-COOLER SYSTEM
MDR report key: 23257729
·
Received October 9, 2025
Report
- Report Number
- 9611109-2025-90419
- Event Type
- Malfunction
- Date Received
- October 9, 2025
- Date of Event
- September 8, 2025
- Report Date
- October 9, 2025
- Manufacturer
- LIVANOVA DEUTSCHLAND
- Product Code
- DWC
- PMA / PMN Number
- N.A.
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
Additional Manufacturer Narrative · 0
A.1.-A.5. THERE WAS NO PATIENT INVOLVEMENT. G.5. THE HEATER-COOLER 16-02-80 IS NOT DISTRIBUTED IN THE USA AND IT IS SIMILAR TO HEATER-COOLER 16-02-85, WHICH IS DISTRIBUTED IN THE USA (510(K) NUMBER: K220635). H11: REPORTEDLY, THE FIELD SERVICE ENGINEER INSPECTED THE UNIT AND FOUND THAT THE CIRCULATION MOTOR WAS NOT WORKING. LIVANOVA INITIATED AN INVESTIGATION. IF ANY ADDITIONAL INFORMATION PERTINENT TO THE REPORTED EVENT IS RECEIVED, IT WILL BE PROVIDED IN A SUPPLEMENTAL REPORT.
Description of Event or Problem · 0
LIVANOVA DEUTSCHLAND RECEIVED A REPORT OF A 3T HEATER/COOLER WHICH GAVE AN ERROR AFTER HAVING SWITCHED OFF THE UNIT. THE EVENT OCCURRED DURING PRIMING. THERE WAS NO PATIENT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2047037 | HEATER-COOLER SYSTEM | CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS | DWC | LIVANOVA DEUTSCHLAND |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |