MECTACER BIOLOX OPTION HEADS
Report
- Report Number
- 3005180920-2025-00980
- Event Type
- Injury
- Date Received
- October 9, 2025
- Date of Event
- September 16, 2025
- Report Date
- October 9, 2025
- Manufacturer
- MEDACTA INTERNATIONAL SA
- Product Code
- LZO
- PMA / PMN Number
- K131518
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- 003
Narratives
BATCH REVIEW PERFORMED ON 23 SEPTEMBER 2025. BALL HEADS: MECTACER 01.29.233H MECTACER HEAD BIOLOX OPTION 12/14 Ø 40 (K131518) LOT 2213136: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 13-SEP-2022. EXPIRATION DATE: 29-AUG-2027. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, ALL ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH ONE SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW. STEM: SMS SOLID 01.36.052 SMS SOLID STEM STD #12 (K181693) LOT 2204985: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 10-AUG-2022. EXPIRATION DATE: 27-JUL-2027. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH ONE SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW. BALL HEADS: MECTACER 01.29.240A MECTACER SLEEVE 12/14 SIZE S (K131518) LOT 2213136: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 13-SEP-2022. EXPIRATION DATE: 29-AUG-2027. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, ALL ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH ONE SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW. CUP: MPACT 01.32.158MH ACETABULAR SHELL Ø58 MULTI-HOLE (K132879) LOT 2245007: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 06-JUN-2023. EXPIRATION DATE: 17-MAY-2028. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH NO SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW LINER: MPACT 01.32.4048HCT FLAT PE HC LINER Ø40/F (K103721) LOT 2239581: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 14-NOV-2022. EXPIRATION DATE: 02-NOV-2027. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH NO SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW CONCLUSIONS: INFECTION IS A KNOWN POSSIBLE COMPLICATION OF ANY SURGERY. ALTHOUGH NO ROOT CAUSE CAN BE ESTABLISHED, THERE IS NO INDICATION THAT ANY ISSUE WITH THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT, AND THE DOCUMENT REVIEW DOES NOT INDICATE ANY POTENTIAL MANUFACTURING RELATED CAUSE.
THE PATIENT CAME IN ABOUT 2 YEARS AFTER THE PRIMARY IN DUE TO INFECTION AND THE CAUSE IS UNKNOWN. EVERYTHING WAS REVISED (HEAD, SLEEVE, CUP LINER AND STEM). THE SURGERY WAS COMPLETED SUCCESSFULLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2481207 | MECTACER BIOLOX OPTION HEADS | MECTACER HEAD BIOLOX OPTION 12/14 Ø 40 | LZO | MEDACTA INTERNATIONAL SA | 01.29.233H | 2213136 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 62 YR | Male | Required Intervention |