FDA Adverse Event Malfunction Summary report: N

CONTACT DETACH

MDR report key: 23253227 · Received October 9, 2025

Report

Report Number
3003442380-2025-14689
Event Type
Malfunction
Date Received
October 9, 2025
Date of Event
September 13, 2025
Report Date
November 4, 2025
Manufacturer
UNOMEDICAL S.A DE C.V
Product Code
FPA
PMA / PMN Number
K041545
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY. COMPLAINT INVESTIGATION RESULTS: A COMPLAINT INVESTIGATION HAS BEEN INITIATED UNDER COMPLAINT INVESTIGATION CHILD RECORD (B)(4). THE BATCH 6011679 IN QUESTION WAS MANUFACTURED AT THE REYNOSA SITE. DEVICE HISTORY RECORD (DHR) REVIEW: THE LOT 6011679 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION (WI) VERSION 100 AND PACKAGING IN THE MACHINE MULTIVAC 14 ON 17-FEB-2025, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY, WELDING OF THE LOT 5B00024 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION (WI) VERSION 34 AND MANUFACTURED IN THE MACHINE LS24-LS25 ON 11-FEB-2025, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY, WELDING OF THE LOT 5B03021 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION (WI) VERSION 34 AND MANUFACTURED IN THE MACHINE LS06-LS07 ON 17-FEB-2025, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY, WELDING OF THE LOT 5B03014 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION (WI) VERSION 34 AND MANUFACTURED IN THE MACHINE LS24-LS25 ON 17-FEB-2025, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY, GLUING CONNECTOR OF THE LOT 5B00041 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION (WI) VERSION 37 AND MANUFACTURED IN THE MACHINE GLUING 3 ON 10-FEB-2025, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY, GLUING CONNECTOR OF THE LOT 5B00046 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION (WI) VERSION 37 AND MANUFACTURED IN THE MACHINE GLUING 3 ON 16-FEB-2025, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY, GLUING CONNECTOR OF THE LOT 4H00372 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION (WI) VERSION 37 AND MANUFACTURED IN THE MACHINE GLUING 3 ON 17-FEB-2025, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY, GLUING CONNECTOR OF THE LOT 5B02913 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION (WI) VERSION 37 AND MANUFACTURED IN THE MACHINE GLUING 3 ON 19-FEB-2025, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY, GLUING CONNECTOR OF THE LOT 5B02914 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION (WI) VERSION 37 AND MANUFACTURED IN THE MACHINE GLUING 3 ON 19-FEB-2025, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY, GLUING CONNECTOR OF THE LOT 5B00046 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION (WI) VERSION 37 AND MANUFACTURED IN THE MACHINE GLUING 3 ON 16-FEB-2025, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DHR SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WAS IDENTIFIED, NOR MAINTENANCE EVENTS WERE RECORDED RELATED TO COMPLAINT CODE. CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: AS A RESULT OF THE FOLLOWING, NO NON-CONFORMANCE (NC) RAISED DURING PRODUCTION RELATED TO COMPLAINT CODE, NO TREND IDENTIFIED FOR THE LOT IN QUESTION, NO FURTHER ACTIONS ARE REQUIRED. THIS COMPLAINT WILL NOT REQUIRE FURTHER ROOT CAUSE INVESTIGATION NOR CORRECTIVE AND PREVENTIVE ACTION (CAPA) PLAN. THEREFORE, THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKET SURVEILLANCE ACTIVITIES.

Description of Event or Problem · 0

REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED STATES. IT WAS REPORTED THAT THE PATIENT FACED TWO INSULIN FLOW BLOCKED ALARM EVENT ON (B)(6) 2025. THE BLOCKAGE WAS AT THE SITE.THE PATIENT REPLACED INFUSION SETS AND RESUMED INSULIN SUCCESSFULLY. NO FURTHER INFORMATION AVAILABLE.

Description of Event or Problem · 0

TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2511690 CONTACT DETACH UNO CONTACT DETACH G29 60/8TCAP 10PK INT FPA UNOMEDICAL S.A DE C.V 1002835 6011679

Patients

Seq Age Sex Outcome Treatment
1 65 YR Female