FDA Adverse Event Injury Summary report: N

INSET II

MDR report key: 23252133 · Received October 9, 2025

Report

Report Number
3003442380-2025-14866
Event Type
Injury
Date Received
October 9, 2025
Date of Event
September 8, 2025
Report Date
October 13, 2025
Manufacturer
UNOMEDICAL DEVICES S.A. DE C.V.
Product Code
FPA
PMA / PMN Number
K032854
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

SUPPLEMENTAL REPORT 01 - (B)(4) - MDR 3003442380-2025-14866. ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY. COMPLAINT INVESTIGATION RESULTS: A COMPLAINT INVESTIGATION HAS BEEN INITIATED UNDER COMPLAINT INVESTIGATION CHILD RECORD (B)(4). THE BATCH 6010351, IN QUESTION WAS MANUFACTURED AT THE REYNOSA SITE. THRESHOLD ANALYSIS: A QUERY WAS RUN ON 09-OCT-2025 AGAINST "FINAL REPORTING DECISION" EQUAL "SERIOUS INJURY AND DEATH", "LOT NUMBER" CRITERIA EQUAL LOT NUMBER "6010351". THE COUNT OF COMPLAINT IS 1 WHICH IS BELOW 3. NO FURTHER STATISTICAL TRENDING ANALYSIS IS REQUIRED. DEVICE HISTORY RECORD (DHR) REVIEW: THE LOT 6010351WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION (WI) VERSION 120 AND MANUFACTURED IN THE LINEA 04 ON 24-NOV-2024, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DEVICE HISTORY RECORD (DHR) SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WAS IDENTIFIED, NOR MAINTENANCE EVENTS WERE RECORDED RELATED TO COMPLAINT CODE. TEST RESULTS: IN ORDER TO TEST THE PRODUCT, NO PHOTO OR PHYSICAL SAMPLE WAS PROVIDED, THEREFORE, THE REFERENCE SAMPLES FROM THE LOT HAVE BEEN REQUESTED. VISUAL TEST ACCORDING TO (WI) VERSION 3 ON REFERENCE SAMPLES, 10 SAMPLES OUT 10 SAMPLES PASSED THE TEST. FUNCTIONAL TEST AIR FLOW ACCORDING TO (WI) VERSION 2 ON REFERENCE SAMPLES, 10 SAMPLES OUT 10 SAMPLES PASSED THE TEST, FOR THE CODE SOFT CANNULA FOUND KINKED UPON REMOVAL FROM INFUSION SITE. CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: AS A RESULT OF THE FOLLOWING: NO PHYSICAL SAMPLES WERE RETURNED. NO NON-CONFORMANCE (NC) RAISED DURING PRODUCTION RELATED TO COMPLAINT CODE, NO THRESHOLDS IDENTIFIED FOR THE LOT IN QUESTION AND SERIOUS INJURY/DEATH, NO FURTHER ACTIONS ARE REQUIRED. THIS IS A COMPLAINT WHICH IS BEING ADDRESSED AS PART OF A CORRECTIVE AND PREVENTIVE ACTION (CAPA) 1768101 AUTOSOFT 90, KINKED SOFT CANNULA ISSUES WHICH HAS BEEN OPENED AS RESULT OF TRENDING ACTIVITIES. THIS IS A COMPLAINT WHICH IS BEING ADDRESSED AS PART OF A CORRECTIVE AND PREVENTIVE ACTION (CAPA) 1768101: "AUTOSOFT 90, KINKED SOFT CANNULA ISSUES WHICH HAS BEEN OPENED AS A RESULT OF TRENDING ACTIVITIES. THIS COMPLAINT IS A REPORTABLE WITH VALID LOT NUMBER/PRODUCT CODE. CORRECTIVE AND PREVENTIVE ACTION (CAPA) DETERMINATION RESULTS: THIS COMPLAINT FALLS UNDER THE SCOPE OF THE CORRECTIVE AND PREVENTIVE ACTION (CAPA) 1768101: "AUTOSOFT 90, KINKED SOFT CANNULA ISSUES" AND WILL COVER INSET I (AUTOSOFT XC DHF-13.8 & 13.13) AND INSET II (AUTOSOFT 90 DHF-14 AND 14.3) PRODUCTS. ROOT CAUSE OF PROBLEM: THE ROOT CAUSE HAS BEEN IDENTIFIED AS: METHOD, MANPOWER, MEASUREMENT, MACHINE: CORRECTIVE ACTION AS A RESULT OF THE INVESTIGATION: THE ACTION PLAN AND DEADLINES IS AS FOLLOWED: THE FOLLOWING ACTIONS WERE ALREADY IMPLEMENTED IN THE NON-CONFORMANCE (NC) · 1. INCLUDE THE DEFECT IN DOCUMENT 4805074 (WORK INSTRUCTION INSET LINE). · 2. IMPROVE GUARDS OF THE CONVEYORS. · 3. UPDATE TRAYS FOR THE STOCK OF CANNULAS. · 4. UPDATE DOCUMENT 3A02003 (QUALITY SPECIFICATION FOR CATHETER FIXTURE FOR SKEWED CATHETERS) FOR INCLUDE THE HANDLING OF THE CATHETER DURING THE PROCESS. · 5. UPDATE THE DOCUMENT 4805074 (WORK INSTRUCTION INSET LINE) TO INCLUDE THE PREVENTIVE ACTIONS IMPLEMENTED IN THE PROCESS (TRAYS). A REMAINING ACTION (TO ADDRESS DESIGN ROOT CAUSE) AND DEADLINES IS AS FOLLOWED: · ADD CYLINDERS TO THE LID TO MAINTAIN IN POSITION THE NEEDLE AND CANNULA DURING TRANSPORTATION AND PRIOR USE (DATABASE 1771074- 30/SEP/2025).

Additional Manufacturer Narrative · 0

INITIAL AND FINAL MDR (B)(4). DEVICE 1 OF 2.

Description of Event or Problem · 0

TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.

Description of Event or Problem · 0

REFERENCE NUMBER (B)(4). EVENT OCCURRED IN AUSTRALIA. IT WAS REPORTED THAT PATIENT EXPERIENCED EVENTS OF INFUSION SET KINKED CANNULA ON (B)(6) 2025. THE EVENT OCCURRED WITHIN THREE HOURS OF INSERTION. THE INSERTION SITE WAS UPPER BUTTOCKS. THE BLOOD GLUCOSE LEVEL WAS 27 MMOL/LAND THE PATIENT WAS TREATED WITH CORRECTION INJECTION VIA MULTIPLE DAILY INJECTION (MDI). PATIENT HAD HIGH/LARGE KETONES AND IT WAS RESOLVED BY REPLACING INFUSION SET AND RESUMED INSULIN SUCCESSFULLY. NO FURTHER INFORMATION AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2508916 INSET II UNO INSET II 60/6 GREY TCAP 10PK INT FPA UNOMEDICAL DEVICES S.A. DE C.V. 1002817 6010351

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown