QUICK SET
Report
- Report Number
- 3003442380-2025-14645
- Event Type
- Malfunction
- Date Received
- October 9, 2025
- Date of Event
- September 11, 2025
- Report Date
- November 1, 2025
- Manufacturer
- UNOMEDICAL DEVICES S.A. DE C.V.
- Product Code
- FPA
- PMA / PMN Number
- K160648
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
E1: PATENT'S CITY: (B)(6). PATIENT'S COUNTRY: (B)(6), U.S.
ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY COMPLAINT INVESTIGATION RESULTS: A COMPLAINT INVESTIGATION HAS BEEN INITIATED UNDER COMPLAINT INVESTIGATION CHILD RECORD (B)(4). THE BATCH 6011992 IN QUESTION WAS MANUFACTURED AT THE REYNOSA SITE. DEVICE HISTORY RECORD (DHR) REVIEW: THE LOT 6011992 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION (WI) VERSION 82 AND MANUFACTURING IN THE MACHINE 12 ON 09-MAR-2025, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY OF THE LOT 5C00035 WAS MANUFACTURED ACCORDING TO THE (WI) VERSION 29 ON 09-MAR-2025, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY OF THE LOT 5C00036 WAS MANUFACTURED ACCORDING TO THE (WI) VERSION 29 ON 09-MAR-2025, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY OF THE LOT 5C00037 WAS MANUFACTURED ACCORDING TO THE (WI) VERSION 29 ON 09-MAR-2025, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DEVICE HISTORY RECORD (DHR) SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WAS IDENTIFIED, NOR MAINTENANCE EVENTS WERE RECORDED RELATED TO COMPLAINT CODE. CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: AS A RESULT OF THE FOLLOWING: NO NON-CONFORMANCE (NC) RAISED DURING PRODUCTION RELATED TO COMPLAINT CODE, NO TREND IDENTIFIED FOR THE LOT IN QUESTION AND MALFUNCTION CODE, NO FURTHER ACTIONS ARE REQUIRED. THIS COMPLAINT WILL NOT REQUIRE FURTHER ROOT CAUSE INVESTIGATION NOR CORRECTIVE AND PREVENTIVE ACTION (CAPA) PLAN. THEREFORE, THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKET SURVEILLANCE ACTIVITIES.
REFERENCE NUMBER (B)(4). EVENT OCCURRED IN (B)(6), UNITED STATES. IT WAS REPORTED THAT THE PATIENT FACED AN EVENT WHERE CANNULA WAS NOT BEEN INSERTED ON (B)(6) 2025 LEADING TO LEAKAGE. THE INFUSION SET WAS USED FOR ONE DAY. THE QUICK RELEASE WAS TIGHTLY CONNECTED. NO FURTHER INFORMATION AVAILABLE.
TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2663303 | QUICK SET | UNO QUICK-SET 60/6 SC1 MECA | FPA | UNOMEDICAL DEVICES S.A. DE C.V. | MMT-399A | 6011992 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |