CONTOUR® PLUS
Report
- Report Number
- 1810909-2025-00152
- Event Type
- Malfunction
- Date Received
- October 8, 2025
- Date of Event
- September 9, 2025
- Report Date
- December 15, 2025
- Manufacturer
- ASCENSIA DIABETES CARE US INC.
- Product Code
- NBW
- PMA / PMN Number
- K191286
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
THE INITIAL REPORTER'S CONTACT DETAILS WERE NOT PROVIDED; THEREFORE, THIS INFORMATION WAS NOT CAPTURED IN SECTION E1. UNKNOWN WAS RECORDED IN SECTION A1, AND NO INFORMATION WAS CAPTURED IN SECTIONS A2, A3 AND A4, AS THE PATIENT'S CREDENTIALS, AGE, SEX AND WEIGHT WERE NOT PROVIDED. THE CONTOUR® PLUS TEST STRIPS USED WITH THE CONTOUR® PLUS METER IS SIMILAR TO THE CONTOUR® NEXT TEST STRIPS USED WITH THE CONTOUR® NEXT METER AVAILABLE IN THE US MARKET. THE CONTOUR® PLUS TEST STRIPS WITH SKU # 7662 AND LOT # 4DQHH07A HAS THE 510K # K191286 AND UDI # (B)(4). THE DEVICE WAS DISTRIBUTED OUTSIDE THE US, THEREFORE, NO GUDID INFORMATION EXISTS.
THE CUSTOMER RETURNED A SUSPECTED CONTOUR® PLUS METER (SERIAL # (B)(6)) AND CONTOUR® PLUS TEST STRIPS FROM LOT 4DQHH07A FOR EVALUATION. IN-HOUSE TESTING WAS PERFORMED WITH THE RETURNED METER AND STRIPS USING BLOOD SPIKED WITH GLUCOSE AT 22.34 MMOL/L AND 3.74 MMOL/L, AS DETERMINED BY A YSI INSTRUMENT, IN FIVE REPLICATES EACH. ALL REPLICATES WERE OUTSIDE OF FIT-FOR-USE LIMITS. ADDITIONALLY, ERROR MESSAGES WERE OBSERVED, INCLUDING E11 (INDICATIVE OF ABNORMAL RESULT), E3 (INDICATIVE OF USED TEST STRIP), AND LO (INDICATIVE OF RESULTS BELOW 0.6 MMOL/L). THE RETURNED TEST STRIPS WERE SENT TO PHC, THE STRIP SUPPLIER, FOR FURTHER EVALUATION. THE ISSUES OF LOW RESULTS AND ERROR MESSAGES WERE REPRODUCED. TESTING WITH RETENTION SAMPLES FROM THE SAME LOT OF TEST STRIPS DEMONSTRATED SATISFACTORY PERFORMANCE, AND REVIEW OF DEVICE HISTORY RECORDS REVEALED NO MANUFACTURING ISSUES. GAS CHROMATOGRAPHY ANALYSIS OF THE RETURNED STRIPS IDENTIFIED CONTAMINATION WITH ETHANOL, PALMITIC ACID, AND SQUALENE. ETHANOL, COMMONLY FOUND IN DISINFECTANTS, IS HIGHLY LIKELY TO HAVE DETERIORATED THE REAGENT, CAUSING LOW RESULTS. PALMITIC ACID AND SQUALENE ARE COMPONENTS OF SEBUM. IF THE CUSTOMER USED HAND SANITIZER AFTER WASHING HANDS AND BEFORE MEASUREMENT AND TOUCHED THE TEST STRIPS OR THE INSIDE OF THE STRIP VIAL BEFORE THE SANITIZER DRIED COMPLETELY, THE REAGENT COULD HAVE DETERIORATED DUE TO EXPOSURE TO THESE SUBSTANCES.
A HEALTHCARE PROFESSIONAL (HCP) FROM CHINA REPORTED THAT A PATIENT'S BLOOD SAMPLE WAS TESTED USING THE CONTOUR® PLUS METER, YIELDING A BLOOD GLUCOSE READING OF 0.8 MMOL/L. THE HCP PERFORMED A SECOND BLOOD GLUCOSE TEST USING AN ALTERNATE CONTOUR® PLUS METER AND THE SAME LOT OF TEST STRIPS, WHICH ALSO PRODUCED A LOW READING. NO SPECIFIC VALUE WAS PROVIDED. SUBSEQUENTLY, LABORATORY TESTING WAS PERFORMED, AND A READING OF 8.2 MMOL/L WAS OBTAINED. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE NURSE WAS ADVISED TO RETURN THE DEVICE FOR EVALUATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2509383 | CONTOUR® PLUS | BLOOD GLUCOSE TEST STRIPS | NBW | ASCENSIA DIABETES CARE US INC. | 7662 | 4DQHH07A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |