INSET II
Report
- Report Number
- 3003442380-2025-14561
- Event Type
- Injury
- Date Received
- October 8, 2025
- Date of Event
- August 10, 2025
- Report Date
- September 24, 2025
- Manufacturer
- UNOMEDICAL DEVICES S.A. DE C.V.
- Product Code
- FPA
- UDI-DI
- 05705244018129
- PMA / PMN Number
- K032854
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- 003
Narratives
INITIAL AND FINAL MDR (B)(4). DEVICE 4 OF 5. ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY: COMPLAINT INVESTIGATION RESULTS: A COMPLAINT INVESTIGATION HAS BEEN INITIATED UNDER COMPLAINT INVESTIGATION CHILD RECORD (B)(4). THE BATCH 6010062 IN QUESTION WAS MANUFACTURED AT THE REYNOSA SITE. THRESHOLD ANALYSIS: A QUERY WAS RUN ON 23-SEP-2025 AGAINST "FINAL REPORTING DECISION EQUAL "SERIOUS INJURY" AND "DEATH", "LOT NUMBER" CRITERIA EQUAL 6010062 THE COUNT OF COMPLAINTS IS 2 WHICH IS BELOW 3. NO FURTHER STATISTICAL TRENDING ANALYSIS IS REQUIRED. DEVICE HISTORY RECORD (DHR) REVIEW: THE LOT 6010062 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION (WI) VERSION 114 AND MANUFACTURED IN THE L175 ON 31-OCT-2024, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DHR SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WAS IDENTIFIED, NOR MAINTENANCE EVENTS WERE RECORDED RELATED TO COMPLAINT CODE. TEST RESULTS: THE REFERENCE SAMPLES WERE ALREADY TESTED IN THE COMPLAINT (B)(4). IN ORDER TO TEST THE PRODUCT, THE RETURNED SAMPLE(S) FROM THE LOT HAVE BEEN REQUESTED. NO PHOTO OR PHYSICAL SAMPLE WAS PROVIDED. CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: AS A RESULT OF THE FOLLOWING: NO PHYSICAL SAMPLES WERE RETURNED. NO DEFECT ON TESTS FOR REFERENCE SAMPLES, NO NON CONFORMANCE (NC) RAISED DURING PRODUCTION UNRELATED TO COMPLAINT CODE, TWO COMPLAINTS IN THRESHOLDS IDENTIFIED FOR THE LOT IN QUESTION AND SERIOUS INJURY/DEATH, NO FURTHER ACTIONS ARE REQUIRED. AS A RESULT OF THE POST MARKET SURVEILLANCE ACTIVITIES, THIS COMPLAINT HAS BEEN EVALUATED AS A TYPE 4 (ESCALATED TO CORRECTIVE AND PREVENTIVE ACTION (CAPA)): THIS IS A COMPLAINT WHICH IS BEING ADDRESSED AS PART OF A CAPA 1768101 AUTOSOFT 90, KINKED SOFT CANNULA ISSUES WHICH HAS BEEN OPENED AS RESULT OF TRENDING ACTIVITIES.
SUPPLEMENTAL REPORT 01 - (B)(4) - MDR 3003442380-2025-14561. CORRECTION: THIS MDR IS BEING SUBMITTED TO CORRECT THE SUBMITTED BRAND NAME UNDER D1, COMMON DEVICE NAME UNDER D2, MODEL NUMBER, SERIAL NUMBER, PRIMARY UNIQUE DEVICE IDENTIFIER (UDI) NUMBER UNDER D4. ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY - COMPLAINT INVESTIGATION RESULTS: A COMPLAINT INVESTIGATION HAS BEEN INITIATED UNDER COMPLAINT INVESTIGATION CHILD RECORD (B)(4). THE BATCH 6010062 IN QUESTION WAS MANUFACTURED AT THE REYNOSA SITE. THRESHOLD ANALYSIS: A QUERY WAS RUN ON 23-SEP-2025 AGAINST "FINAL REPORTING DECISION EQUAL "SERIOUS INJURY" AND "DEATH", "LOT NUMBER" CRITERIA EQUAL 6010062 THE COUNT OF COMPLAINTS IS 2 WHICH IS BELOW 3. NO FURTHER STATISTICAL TRENDING ANALYSIS IS REQUIRED. DEVICE HISTORY RECORD (DHR) REVIEW: THE LOT 6010062 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION (WI) VERSION 114 AND MANUFACTURED IN THE L175 ON 31-OCT-2024, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DHR SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WAS IDENTIFIED, NOR MAINTENANCE EVENTS WERE RECORDED RELATED TO COMPLAINT CODE. TEST RESULTS: THE REFERENCE SAMPLES WERE ALREADY TESTED IN THE COMPLAINT (B)(4). ALL TEST RESULTS WERE WITHIN SPECIFICATIONS AS PER THE TEST REPORT (B)(4). IN ORDER TO TEST THE PRODUCT, THE RETURNED SAMPLE(S) FROM THE LOT HAVE BEEN REQUESTED. NO PHOTO OR PHYSICAL SAMPLE WAS PROVIDED. CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: AS A RESULT OF THE FOLLOWING: NO PHYSICAL SAMPLES WERE RETURNED. NO DEFECT ON TESTS FOR REFERENCE SAMPLES, NO NON-CONFORMANCE (NC) RAISED DURING PRODUCTION UNRELATED TO COMPLAINT CODE, TWO COMPLAINTS IN THRESHOLDS IDENTIFIED FOR THE LOT IN QUESTION AND SERIOUS INJURY/DEATH, NO FURTHER ACTIONS ARE REQUIRED. AS A RESULT OF THE POST MARKET SURVEILLANCE ACTIVITIES, THIS COMPLAINT HAS BEEN EVALUATED AS A TYPE 4 (ESCALATED TO CAPA): THIS IS A COMPLAINT WHICH IS BEING ADDRESSED AS PART OF A CORRECTIVE AND PREVENTIVE ACTION (CAPA) 1768101 AUTOSOFT 90, KINKED SOFT CANNULA ISSUES WHICH HAS BEEN OPENED AS RESULT OF TRENDING ACTIVITIES.
REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED STATES. IT WAS REPORTED THAT THE PATIENT WENT TO THE EMERGENCY ROOM (ER) AND EVENTUALLY GOT HOSPITALIZED ON (B)(6) 2025 DUE TO HYPERGLYCEMIA. THE PATIENT REPORTED A BENT CANNULA. THE BLOOD GLUCOSE LEVEL WAS 500 MG/DL AT THE TIME OF EVENT AND THE PATIENT GOT TREATED WITH INTRAVENOUS FLUIDS OF SALINE AND INSULIN. THE PATIENT WAS RELEASED TWO DAYS AFTER BEING HOSPITALIZED. CUSTOMER REPLACED THE INFUSION SET AND RESUMED INSULIN DELIVERY SUCCESSFULLY. NO FURTHER INFORMATION AVAILABLE.
TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2574516 | INSET II | UNO INSET II 60/6 GREY TCAP 10PK INT | FPA | UNOMEDICAL DEVICES S.A. DE C.V. | 1002817 | 6010062 | 05705244018129 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 67 YR | Male | Required Intervention| H |