GRASPING FORCEPS, SHARK TEETH, 5 FR., SEMIFLEXIBLE

MDR report key: 23237655 · Received October 7, 2025

Report

Report Number
9610773-2025-06228
Event Type
Malfunction
Date Received
October 7, 2025
Date of Event
September 8, 2025
Report Date
October 7, 2025
Manufacturer
OLYMPUS WINTER & IBE GMBH
Product Code
NWW
PMA / PMN Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE WAS RETURNED TO OLYMPUS FOR INSPECTION. A ROOT CAUSE COULD NOT BE IDENTIFIED. BASED ON THE RESULTS OF THE INVESTIGATION, IT IS LIKELY THE FOLLOWING LED TO THE MALFUNCTION: WIRE (SHAFT) THAT PUSHES AND PULLS THE FORCEPS PORT (JAWS) WAS FRACTURED AT THE CONNECTING SECTION TO THE HANDLE. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.

Description of Event or Problem · 0

IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SUBJECT DEVICE EXHIBITED THE WIRE (SHAFT) THAT PUSHES AND PULLS THE FORCEPS PORT (JAWS) WAS FRACTURED AT THE CONNECTING SECTION TO THE HANDLE. THE SPRING INSIDE THE HANDLE IS DEFORMED. THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2746934 GRASPING FORCEPS, SHARK TEETH, 5 FR., SEMIFLEXIBLE FLEXIBLE ENDOSCOPIC GRASPING FORCEPS, REUSABLE NWW OLYMPUS WINTER & IBE GMBH A4825

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown