FDA Adverse Event
Malfunction
Summary report: N
GRASPING FORCEPS, SHARK TEETH, 5 FR., SEMIFLEXIBLE
MDR report key: 23237655
·
Received October 7, 2025
Report
- Report Number
- 9610773-2025-06228
- Event Type
- Malfunction
- Date Received
- October 7, 2025
- Date of Event
- September 8, 2025
- Report Date
- October 7, 2025
- Manufacturer
- OLYMPUS WINTER & IBE GMBH
- Product Code
- NWW
- PMA / PMN Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- 003
Narratives
Additional Manufacturer Narrative · 0
THE DEVICE WAS RETURNED TO OLYMPUS FOR INSPECTION. A ROOT CAUSE COULD NOT BE IDENTIFIED. BASED ON THE RESULTS OF THE INVESTIGATION, IT IS LIKELY THE FOLLOWING LED TO THE MALFUNCTION: WIRE (SHAFT) THAT PUSHES AND PULLS THE FORCEPS PORT (JAWS) WAS FRACTURED AT THE CONNECTING SECTION TO THE HANDLE. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.
Description of Event or Problem · 0
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SUBJECT DEVICE EXHIBITED THE WIRE (SHAFT) THAT PUSHES AND PULLS THE FORCEPS PORT (JAWS) WAS FRACTURED AT THE CONNECTING SECTION TO THE HANDLE. THE SPRING INSIDE THE HANDLE IS DEFORMED. THERE WAS NO PATIENT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2746934 | GRASPING FORCEPS, SHARK TEETH, 5 FR., SEMIFLEXIBLE | FLEXIBLE ENDOSCOPIC GRASPING FORCEPS, REUSABLE | NWW | OLYMPUS WINTER & IBE GMBH | A4825 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |