BUR,RND,CLEARCUT,8 FLUT,5.5MMX 13CM
Report
- Report Number
- 1220246-2025-04483
- Event Type
- Malfunction
- Date Received
- October 7, 2025
- Date of Event
- July 2, 2025
- Report Date
- October 7, 2025
- Manufacturer
- ARTHREX, INC.
- Product Code
- GFF
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- 501
Narratives
THE COMPLAINT ALLEGATION WAS CONFIRMED. ONE UNPACKAGED AR-8550CRE BATCH 15326989 WAS RECEIVED FOR INVESTIGATION. VISUAL INSPECTION REVEALED THAT THE MOLDED CLEAR TIP WAS NOT ASSEMBLED ONTO THE SHAFT TIP, MOST LIKELY DUE TO BREAKAGE. SCRATCHES AND DENTS WERE OBSERVED ON THE CUTTING TIP OF THE OUTER SHAFT, LIKELY CAUSED BY EXCESSIVE FORCE. FUNCTIONAL TESTING WAS PERFORMED BY ROTATING THE INNER AND OUTER SHAFTS; NO RESISTANCE WAS NOTED. THE INNER SHAFT WAS REMOVED, AND BROWN MARKS WERE OBSERVED, POSSIBLY INDICATING INSUFFICIENT IRRIGATION. THE MOST LIKELY REASON FOR THE REPORTED FAILURE IS USER ERROR, INCLUDING EXCESSIVE SIDE LOADING ON THE DEVICE DURING USE. DIRECTIONS FOR USE. DFU-0215-EOR1_FMT_EN. DISPOSABLE ARTHROSCOPIC SHAVER BLADES, BURRS, POWERPICK¿, POWERASP¿, NANORESECTION¿, AND SHAVERDRILL¿ DEVICES. F. PRECAUTIONS, 7. EXCESSIVE "SIDE-LOADING" ON THE DEVICE DURING USE DOES NOT IMPROVE CUTTING PERFORMANCE AND IN EXTREME CASES WILL CAUSE IRREVERSIBLE DAMAGE TO THE DEVICE.
IT WAS REPORTED THAT DURING A SURGERY THE BURR GOT STUCK IN THE GENERATOR. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH THE SAME DEVICE. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY. UPDATE (B)(6) 24-SEP-2025: THE DEVICE WAS RETURNED TO ARTHREX INC FOR INVESTIGATION. DURING INVESTIGATION IT WAS NOTED THAT THE CLEAR TIP OF THE RECEIVED DEVICE WAS BROKEN OFF.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1487539 | BUR,RND,CLEARCUT,8 FLUT,5.5MMX 13CM | BURR,SURGICAL | GFF | ARTHREX, INC. | BUR,RND,CLEARCUT,8 FLUT,5.5MMX 13CM | 15326989 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |