FDA Adverse Event Injury Summary report: N

S8 AUTOSET SPIRIT II UK

MDR report key: 2323482 · Received October 28, 2011

Report

Report Number
3004604967-2011-00054
Event Type
Injury
Date Received
October 28, 2011
Date of Event
September 1, 2011
Report Date
October 28, 2011
Manufacturer
RESMED LTD
Product Code
BZD
PMA / PMN Number
K071171
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PT DID NOT SEEK MEDICAL ATTENTION. RESMED LTD PERFORMED AN ENGINEERING INVESTIGATION FOUND THE METAL CRIMPING PIN THAT HOLDS THE CONDUCTIVE PORTION OF THE POWER CORD WIRE IN PLACE WAS BROKEN OFF. THIS CAUSED POOR ELECTRICAL CONTACT WHICH LED TO ARCING THAT OVERHEATED THE AC CONNECTOR'S INSULATION. THE MOST LIKELY CAUSE OF THE BROKEN PIN OF THE POWER CORD AC CONNECTOR WAS A BENDING FORCE APPLIED AT THE JOIN OF THE CRIMPING PIN AND THE WIRE WHICH APPEARS TO BE THE MOST VULNERABLE SITE IN THIS POWER CORD.

Description of Event or Problem · 1

IT WAS REPORTED TO RESMED (B)(4) THAT AN S8 AUTOSET SPIRIT II DEVICE WAS SMOKING AT THE POWER CORD. AROUND 6AM, THE PT WAS WOKEN UP WHEN THEY REPORTEDLY HAD DIFFICULTY BREATHING AND FOUND SPARKS AND SMOKE COMING FROM THEIR DEVICE WHERE THE POWER CORD CONNECTS TO THE MACHINE. THE PT MANAGED TO UNPLUG THE MACHINE FROM THE MAIN POWER SUPPLY. THE PT ALLEGES SHE SOON COLLAPSED THEREAFTER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 S8 AUTOSET SPIRIT II UK BZD RESMED LTD 33146

Patients

Seq Age Sex Outcome Treatment
1 Other