FDA Adverse Event Injury Summary report: N

CAPIOX FX25 OXYGENATOR

MDR report key: 23230593 · Received October 6, 2025

Report

Report Number
9681834-2025-00181
Event Type
Injury
Date Received
October 6, 2025
Date of Event
September 8, 2025
Report Date
October 6, 2025
Manufacturer
TERUMO CORPORATION, ASHITAKA
Product Code
DTZ
PMA / PMN Number
K071494
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

D6A: IMPLANTED DATE: DEVICE WAS NOT IMPLANTED. D6B: EXPLANTED DATE: DEVICE WAS NOT EXPLANTED. E3: OCCUPATION: UNKNOWN. G4: 510(K) NO.: K130520. THE MANUFACTURING RECORD AND THE SHIPPING INSPECTION RECORD OF THE ACTUAL DEVICE NO ANOMALY WAS FOUND. PAST COMPLAINT FILE NO OTHER SIMILAR REPORT OF THE PRODUCT WITH THE INVOLVED PRODUCT CODE/LOT NUMBER WAS FOUND. BASED ON THE INVESTIGATION RESULT, NO ANOMALY WAS FOUND IN THE MANUFACTURING RECORD AND THE SHIPPING INSPECTION RECORD. SINCE THE ACTUAL DEVICE WAS NOT RETURNED AND DETAILED INFORMATION WAS NOT AVAILABLE, IT WAS NOT POSSIBLE TO CLARIFY THE CAUSE OF OCCURRENCE. TERUMO MEDICAL CORPORATION (TMC) (IMPORTER) REGISTRATION NO. 2243441 IS SUBMITTING THIS REPORT ON BEHALF OF ASHITAKA FACTORY OF TERUMO CORPORATION (MANUFACTURER) REGISTRATION NO. 9681834.

Description of Event or Problem · 0

TERUMO MEDICAL CORPORATION RECEIVED THE FOLLOWING REPORTED INFORMATION: DURING A BENTALL PROCEDURE, FOAM FORMATION IN THE VENOUS RESERVOIR AFTER 5 HOURS OF CBP TIME AND LEADS TO CHANGE THE OXY WITH ANOTHER FX25 OXYGENATOR. THE PROCEDURE OUTCOME WAS NOT REPORTED. THERE WAS NO HARM TO THE PATIENT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2746201 CAPIOX FX25 OXYGENATOR OXYGENATOR, CARDIOPULMONARY BYPASS DTZ TERUMO CORPORATION, ASHITAKA CX*FX25RW 240917

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Required Intervention