ANN BLUNT TIP SCREW 4X42MM
Report
- Report Number
- 0009613350-2025-00755
- Event Type
- Malfunction
- Date Received
- October 6, 2025
- Date of Event
- September 8, 2025
- Report Date
- February 20, 2026
- Manufacturer
- ZIMMER GMBH
- Product Code
- HSB
- PMA / PMN Number
- K200814
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
(B)(4). D10: ITEM # 47249616109, PROXIMAL HUMERUS, LEFT, ÿ 9X160MM, LOT # 3233558. ITEM # 47248603840, ANN BLUNT TIP SCREW 4X38MM, LOT # 3237314. ITEM # 47248604040, ANN BLUNT TIP SCREW 4X40MM, LOT # 3242834. ITEM # 47248605040, ANN BLUNT TIP SCREW 4X50MM, LOT # 3218771. ITEM # 47248612440, ANN CORT BONE SCREW 4 X 24MM, LOT # 3216555. ITEM # 47248612640, ANN CORT BONE SCREW 4 X 26MM, LOT # 3228725. ITEM # 47248613640, ANN CORT BONE SCREW 4 X 36MM, LOT # 3189224. ITEM # 47248801000, AFFIXUS PH NL CAP 0MM, LOT # 3238877. ITEM # 47249616109, ANN PH NAIL LT 9X160MM, LOT # 3233558. G2: REPORT SOURCE JAPAN. THE DEVICE WILL NOT BE RETURNED FOR ANALYSIS; HOWEVER, AN INVESTIGATION OF THE REPORTED EVENT IS IN PROGRESS. ONCE THE INVESTIGATION IS COMPLETED, A SUPPLEMENTAL MEDWATCH 3500A WILL BE SUBMITTED.
(B)(4). THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION. THE FOLLOWING SECTIONS WERE UPDATED: B4, G3, G6, H2, H3, H6, H11. NO PRODUCT WAS RETURNED OR PICTURES PROVIDED; A PRODUCT EVALUATION COULD NOT BE PERFORMED. A REVIEW OF THE DEVICE MANUFACTURING RECORDS CONFIRMED NO ABNORMALITIES OR DEVIATIONS. A DEFINITIVE ROOT CAUSE COULD NOT BE DETERMINED. HOWEVER BASED ON THE INVESTIGATION IT COULD BE ASSUMED THAT POSSIBLE CONTRIBUTING FACTORS TO THE MIGRATION OF THE SCREW MIGHT BE MULTIFACTORIAL RELATED TO EITHER PATIENT CONDITION, BEHAVIOR OR IMPLANTATION PROCEDURE. IF AND TO WHAT EXTENT ANY OF THESE ASPECTS MAY HAVE INFLUENCED THE MIGRATION OF THE SCREW REMAINS UNKNOWN. IN CONCLUSION, AS THE CAUSE MAY BE MULTIFACTORIAL AN EXACT ROOT CAUSE COULD NOT BE IDENTIFIED FOR THE MIGRATION OF THE SCREW. AN ADDITIONAL DEEPER INVESTIGATION WAS PERFORMED WHICH IDENTIFIED THE DESIGN LIMITATION OF THE CORELOCK MECHANISM AS A POTENTIAL CONTRIBUTING FACTOR. HOWEVER, AS FURTHER BIOMECHANICAL TESTING WAS CARRIED OUT, AND THE PERFORMANCE IS IN AN ACCEPTABLE RANGE IN COMPARISON WITH LEGALLY MARKETED SIMILAR DEVICES, NO DESIGN CHANGES WERE CONDUCTED. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.
IT WAS REPORTED THAT 1 MONTH POST-IMPLANTATION, SURGEON FOUND ALL OF THE PROXIMAL SCREWS WAS BACKED OUT FROM THE PROPER POSITION. THE SURGEON KEEPS AN EYE ON THE PATIENT CONDITION AS WELL AS NO REVISION WILL BE PLANNED SO FAR. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2343726 | ANN BLUNT TIP SCREW 4X42MM | TRAUMA, PLATE | HSB | ZIMMER GMBH | 3208883 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Female | SEE H11 NARRATIVE |