Narratives

Additional Manufacturer Narrative · 0

AS A RESULT OF THIS MALFUNCTION, THE POTENTIAL FOR SURGICAL INTERVENTION EXISTS TO PRECLUDE PERMANENT DAMAGE TO A BODY STRUCTURE OR PERMANENT IMPAIRMENT OF A BODY FUNCTION AS EVIDENCED BY PREVIOUS REPORTED EVENTS WITH SIMILAR FILES. THIS EVENT, THEREFORE, IS REPORTABLE PER 21CFR PART 803. THE DEVICE WAS NOT RETURNED FOR EVALUATION. HOWEVER, THE LOT NUMBER WAS PROVIDED AND RETAINED-PRODUCT TESTING AND/OR DHR REVIEW ARE PLANNED. THE RESULTS WILL BE SUBMITTED AS THEY BECOME AVAILABLE.

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION RECEIVED: WE RECEIVED THE FOLLOWING INFORMATION REGARDING THIS COMPLAINT: THE BROKEN PART HAS BEEN REMOVED. NO FURTHER TREATMENT. IT'S A NEWLY OPENED FILE. THE PRODUCT INVOLVED IS UNAVAILABLE, NOT RETURN. THIS IS A FOLLOW UP REPORT FOR THIS ADDITIONAL INFORMATION. INVESTIGATION RESULTS: INVOLVED PRODUCT THAT BROKE DURING USE WAS NOT RETURNED AND CANNOT BE ANALYZED. MOREOVER, NO UNUSED FILE IS AVAILABLE FOR EVALUATION. NOTHING UNUSUAL TO REPORT WAS FOUND DURING DHR REVIEW (BATCH #1880056). ROOT CAUSES ARE NOT IDENTIFIED. WE WILL TRACK THIS KIND OF EVENT AND MONITOR THE TREND. POTENTIAL ROOT CAUSES MAY BE INCORRECT TECHNIQUE, OVERUSE (USING IN A CANAL TOO CURVED AS IT COULD BE THE CASE HERE), EXCESSIVE WEAR, PATIENT CONDITION AND BEHAVIOR DURING TREATMENT OR MATERIAL ISSUE (NO ANALYSIS OF THE BROKEN FRAGMENTS POSSIBLE). IN ADDITION, THERE ARE FURTHER CIRCUMSTANCES ON THE CONDITIONS IN DENTISTRY (USED DISINFECTANTS, TRAINING/KNOWLEDGE STATUS), OR ANY OTHER ENVIRONMENTAL CONDITIONS, WHICH ARE UNKNOWN TO US AND MAY ALSO HAVE AN IMPACT ON THE REPORTED FAILURE MODE.

Description of Event or Problem · 0

IN THIS EVENT IT IS REPORTED THAT A WAVEONE GOLD PRIMARY 6-FILE STER 25MM FILE BROKE DURING USE. REPORTEDLY, NO INJURY BUT INFORMATION ABOUT IF BROKEN PART RETRIEVED OR NOT IS UNKNOWN. FURTHER INFORMATION REQUESTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2263726 WAVEONE GOLD PRIMARY 6-FILE STER 25MM FILE, PULP CANAL, ENDODONTIC EKS MAILLEFER INSTRUMENTS HOLDING SARL 1880056

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown