DETERMINE HIV-1/2 AG/AB COMBO 25T
Report
- Report Number
- 1221359-2025-00413
- Event Type
- Malfunction
- Date Received
- October 3, 2025
- Date of Event
- May 6, 2025
- Report Date
- October 3, 2025
- Manufacturer
- ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
- Product Code
- MZF
- PMA / PMN Number
- BP120037
- Removal / Correction Number
- 1221359-07/31/2025-002-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NM, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT 0000950116 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 7D2648 LOT 0000950116 AND DEVICE PART NUMBER 10732998 LOT 949678. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE POSITIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT 0000950116 SHOWED THAT THE COMPLAINT RATE IS 0.0298%. AS PART OF AN INVESTIGATION INTO PRELIMINARY FALSE REACTIVE RESULTS, ABBOTT FOUND THAT FOR KIT LOT 0000950116, THE ROOT CAUSE OF THE RESULTS WAS DUE TO A SUBCOMPONENT USED IN THE MANUFACTURE OF DETERMINE HIV-1/2 AG/AG COMBO. THE ISSUE HAS BEEN ISOLATED TO SPECIFIC LOTS OF THE SUBCOMPONENT, WHICH WERE USED TO MANUFACTURE SPECIFIC TEST KIT LOTS. THE NECESSARY ACTIONS TO PREVENT RECURRENCE HAVE BEEN TAKEN, AND THE CORRECTION OF PRODUCT IN THE FIELD WAS CONDUCTED.
THE CUSTOMER REPORTED TWO (2) FALSE POSITIVE RESULTS WITH THE DETERMINE HIV 1/2 AB/AG COMBO TEST WITH A BLOOD SAMPLE PERFORMED ON (B)(6) 2025 AND (B)(6) 2025, RESPECTIVELY. THIS REPORT ADDRESSES PATIENT ONE (1) OF TWO (2). THE CUSTOMER RECEIVED A POSITIVE RESULT WITH THE DETERMINE HIV 1/2 AB/AG COMBO TEST ON (B)(6) 2025, AND A CONFIRMATORY TEST (GEN 4 HIV 1/2 AB/AG GENIUS COMBO) WAS PERFORMED ON THE SAME DAY, WHICH GENERATED A NEGATIVE RESULT. ALTHOUGH REQUESTED, NO INFORMATION ABOUT THE PATIENT, THEIR TREATMENT, OR OUTCOME WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1588522 | DETERMINE HIV-1/2 AG/AB COMBO 25T | TEST, HIV DETECTION | MZF | ABBOTT DIAGNOSTICS SCARBOROUGH, INC. | 0000950116 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 27 YR | Female |