INSPIRA SMOOTH COHESIVE SILICONE GEL FILLED BREAST IMPLANT
Report
- Report Number
- 9617229-2025-16851
- Event Type
- Injury
- Date Received
- October 3, 2025
- Date of Event
- August 5, 2025
- Report Date
- January 14, 2026
- Manufacturer
- ALLERGAN (COSTA RICA)
- Product Code
- FTR
- PMA / PMN Number
- P020056
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
LABORATORY ANALYSIS SUMMARY: THE DEVICE RELATED TO THE REPORTED EVENT OF RUPTURE WAS RECEIVED ON DECEMBER 16, 2025, WITH LOT NUMBER 2961188. PER THE INVESTIGATION PROCEDURE, THE DEVICE IS ANALYZED THROUGH VISUAL INSPECTION MICROSCOPIC INSPECTION IF OPENINGS ARE OBSERVED AND A WEIGHT VERIFICATION. PER THE ANALYSIS PERFORMED, THE ASSESSMENTS OF THE COMPLAINTS AND ANY POTENTIAL MANUFACTURING ISSUE ARE DISPLAYED ALONG WITH ANY FURTHER ACTIONS REQUIRED: ¿ RUPTURE: OBSERVED BROKEN DEVICE THROUGH MICROSCOPIC INSPECTION ASSESSED AS UNIDENTIFIED (TEAR) OPENING. NO FURTHER ACTIONS ARE REQUIRED AS IT IS NOT RELATED TO THE MANUFACTURING PROCESS. NO ADDITIONAL OBSERVATIONS PERFORMED ON THE DEVICE. NO FURTHER ACTIONS ARE REQUIRED.
A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. REASON FOR REOPERATION: RUPTURE.
HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE ¿RUPTURED IMPLANT¿ DIAGNOSED VIA MRI. DEVICE WAS EXPLANTED AND REPLACED.
HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE ¿RUPTURED IMPLANT¿ DIAGNOSED VIA MRI. DEVICE WAS EXPLANTED AND REPLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2433594 | INSPIRA SMOOTH COHESIVE SILICONE GEL FILLED BREAST IMPLANT | PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED | FTR | ALLERGAN (COSTA RICA) | 2961188 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 55 YR | Female | Required Intervention |