MASSCHECK® DRIED BLOOD SPOT CONTROL BI-LEVEL (I+II)
Report
- Report Number
- 3003951180-2025-00001
- Event Type
- Malfunction
- Date Received
- October 3, 2025
- Manufacturer
- CHROMSYSTEMS INSTRUMENTS & CHEMICALS GMBH
- Product Code
- JJY
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
A FOREIGN CUSTOMER (LOCATED IN GERMANY) REPORTED THE ISSUE INITIALLY. EUROPEAN GDPR RESTRICTS INCLUDING PERSONAL DETAILS. FOR REFERENCE AND FOR FDA FOLLOW-UP WE PROVIDE OUR CHROMSYSTEMS COMPANY RA CONTACT: DR. (B)(6), [email protected], CHROMSYSTEMS INSTRUMENTS & CHEMICALS GMBH, AM HAAG 12, 82166 GRÄFELFING.
IN THE FOLLOWING WE DESCRIBE THE ISSUE REGARDING THE MASSCHECK® DRIED BLOOD SPOT CONTROL BI-LEVEL (I+II)(WITH UDI: (B)(4)(ORDER NO. (B)(4), WHICH ITSELF CONSISTS OF MASSCHECK® DRIED BLOOD SPOT CONTROL LEVEL I (ORDER NO. (B)(4) AND MASSCHECK® DRIED BLOOD SPOT CONTROL LEVEL II (ORDER NO. (B)(4). WE FOUND THAT THE TARGET CONCENTRATION OF PHENYLALANINE INDICATED IN THE LEAFLET, DETERMINED USING THE NON-DERIVATISED METHODS 57000, 57000/F, 57111 AND 57111/F IS TOO HIGH FOR DRIED BLOOD SPOT CONTROL LEVEL I (ORDER NO. (B)(4) AND TOO LOW FOR DRIED BLOOD SPOT CONTROL LEVEL II (ORDER NO. (B)(4). THE CONTROLS ARE PRIMARILY USED TO VERIFY THE ACCURACY AND PRECISION OF THE ANALYTICAL PROCEDURE. IF THE MEASURED VALUES ARE OUTSIDE THE RANGES SPECIFIED ON THE CONTROL PACKAGE INSERT, THE MEASUREMENT SYSTEM, THE SAMPLE PREPARATION PROCEDURE AND THE CALCULATION OF THE ANALYSIS VALUES MUST BE CHECKED. THIS MAY RESULT IN A DELAY OR FAILURE TO TRANSMIT SCREENING RESULTS. FURTHERMORE, THE CONTROLS CAN BE USED TO INTRODUCE OR ADJUST A FACTORISATION (RELATIVE RESPONSE FACTOR, RRF). THIS IS USUALLY USED TO HARMONISE MEASURED VALUES ON DEVICES OVER TIME AND ON DIFFERENT DEVICES AND TO ENABLE THE USE OF A UNIFORM CUT-OFF VALUE. FACTORING IN INCORRECT VALUES ADJUSTS NOT ONLY THE CONTROLS BUT ALSO ALL PATIENT SAMPLES BY THE SAME FACTOR. SINCE THE CUT-OFF VALUE IS NOT ADJUSTED ACCORDINGLY, THERE IS A RISK THAT METABOLIC DISORDERS WILL NOT BE DETECTED IN NEWBORN SCREENING DUE TO INCORRECT FACTORING. IN THE SCENARIOS DESCRIBED, PATIENTS WITH A DISEASE FOR WHICH PHENYLALANINE IS USED AS A MARKER IN NEWBORN SCREENING CAN LEAD TO A CRITICAL CONDITION WITH SERIOUS HEALTH CONSEQUENCES IF THE METABOLIC DISEASE IS NOT DETECTED OR NOT DETECTED IN TIME.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2427404 | MASSCHECK® DRIED BLOOD SPOT CONTROL BI-LEVEL (I+II) | DRIED BLOOD SPOT CONTROL BI-LEVEL (I+II) | JJY | CHROMSYSTEMS INSTRUMENTS & CHEMICALS GMBH | 0191 | 4924, 1425 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |