FDA Adverse Event Injury Summary report: N

INSET

MDR report key: 23206956 · Received October 3, 2025

Report

Report Number
3003442380-2025-14382
Event Type
Injury
Date Received
October 3, 2025
Date of Event
July 27, 2025
Report Date
September 17, 2025
Manufacturer
UNOMEDICAL DEVICES S.A. DE C.V.
Product Code
FPA
PMA / PMN Number
K032854
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

INITIAL AND FINAL MDR (B)(4). DEVICE 1 OF 4.

Additional Manufacturer Narrative · 0

SUPPLEMENTAL REPORT 01 - (B)(4) - MDR 3003442380-2025-14382. ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY: COMPLAINT INVESTIGATION RESULTS: A COMPLAINT INVESTIGATION HAS BEEN INITIATED UNDER COMPLAINT INVESTIGATION CHILD RECORD (B)(4). THE BATCH 6010219, IN QUESTION WAS MANUFACTURED AT THE REYNOSA SITE. THRESHOLD ANALYSIS: A QUERY WAS RUN ON 16-SEP-2025 AGAINST "FINAL REPORTING DECISION EQUAL "SERIOUS INJURY" AND "DEATH", "LOT NUMBER" CRITERIA EQUAL 6010219. THE COUNT OF COMPLAINT IS 1 WHICH IS BELOW 3. NO FURTHER STATISTICAL TRENDING ANALYSIS IS REQUIRED. DEVICE HISTORY RECORD (DHR) REVIEW: THE LOT 6010219 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION (WI) VERSION 120 AND MANUFACTURED IN LINE L192 ON 14-NOV-2024, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DHR SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WAS IDENTIFIED, NOR MAINTENANCE EVENTS WERE RECORDED RELATED TO COMPLAINT CODE. TEST RESULTS: NO PHOTO WAS PROVIDED. IN ORDER TO TEST THE PRODUCT, THE RETURNED SAMPLE FROM THE LOT HAVE BEEN REQUESTED. NO RETURNED SAMPLES. CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: AS A RESULT OF THE FOLLOWING: NO PHYSICAL SAMPLES WERE RETURNED. NO NON-CONFORMANCE (NC) RAISED DURING PRODUCTION RELATED TO COMPLAINT CODE, ONE COMPLAINT IN THRESHOLDS IDENTIFIED FOR THE LOT IN QUESTION AND SERIOUS INJURY/DEATH, NO FURTHER ACTIONS ARE REQUIRED. THIS IS A COMPLAINT WHICH IS BEING ADDRESSED AS PART OF A CORRECTIVE AND PREVENTIVE ACTION (CAPA) 1768101 AUTOSOFT 90, KINKED SOFT CANNULA ISSUES WHICH HAS BEEN OPENED AS RESULT OF TRENDING ACTIVITIES. THIS IS A COMPLAINT WHICH IS BEING ADDRESSED AS PART OF A CAPA 1768101: "AUTOSOFT 90, KINKED SOFT CANNULA ISSUES WHICH HAS BEEN OPENED AS A RESULT OF TRENDING ACTIVITIES. THIS COMPLAINT IS A REPORTABLE WITH VALID LOT NUMBER/PRODUCT CODE. CAPA DETERMINATION RESULTS: THIS COMPLAINT FALLS UNDER THE SCOPE OF THE CAPA 1768101: "AUTOSOFT 90, KINKED SOFT CANNULA ISSUES" AND WILL COVER INSET I (AUTOSOFT XC DHF-13.8 & 13.13) AND INSET II (AUTOSOFT 90 DHF-14 AND 14.3) PRODUCTS. ROOT CAUSE OF PROBLEM: THE ROOT CAUSE HAS BEEN IDENTIFIED AS: METHOD, MANPOWER, MEASUREMENT, MACHINE: CORRECTIVE ACTION AS A RESULT OF THE INVESTIGATION: THE ACTION PLAN AND DEADLINES IS AS FOLLOWED: THE FOLLOWING ACTIONS WERE ALREADY IMPLEMENTED IN THE NC. 1. INCLUDE THE DEFECT IN DOCUMENT 4805074 (WORK INSTRUCTION INSET LINE). 2. IMPROVE GUARDS OF THE CONVEYORS. 3. UPDATE TRAYS FOR THE STOCK OF CANNULAS. 4. UPDATE DOCUMENT 3A02003 (QUALITY SPECIFICATION FOR CATHETER FIXTURE FOR SKEWED CATHETERS) FOR INCLUDE THE HANDLING OF THE CATHETER DURING THE PROCESS. 5. UPDATE THE DOCUMENT 4805074 (WORK INSTRUCTION INSET LINE) TO INCLUDE THE PREVENTIVE ACTIONS IMPLEMENTED IN THE PROCESS (TRAYS). A REMAINING ACTION (TO ADDRESS DESIGN ROOT CAUSE) AND DEADLINES IS AS FOLLOWED: ADD CYLINDERS TO THE LID TO MAINTAIN IN POSITION THE NEEDLE AND CANNULA DURING TRANSPORTATION AND PRIOR USE (DATABASE 1771074- 30/SEP/2025).

Description of Event or Problem · 0

REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED STATES. IT WAS REPORTED THAT THE PATIENT WENT TO THE EMERGENCY ROOM (ER) ON (B)(6) 2025 DUE TO HYPERGLYCEMIA. THE DURATION OF EMERGENCY ROOM (ER) STAY WAS SEVEN HOURS. THE BLOOD GLUCOSE LEVEL EXCEEDED 500 MG/DL AND THE PATIENT GOT TREATED WITH INTRAVENOUS FLUIDS OF SALINE AND MANUAL INJECTION OF INSULIN. THE PATIENT REPORTED INFUSION SET CANNULA WAS KINKED. THE PATIENT REPLACED THE INFUSION SET AND RESUMED INSULIN DELIVERIES SUCCESSFULLY. NO FURTHER INFORMATION AVAILABLE.

Description of Event or Problem · 0

TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
292492 INSET UNO INSET I 60/6 GREY TCAP 10PK INT FPA UNOMEDICAL DEVICES S.A. DE C.V. 1001680 6010219

Patients

Seq Age Sex Outcome Treatment
1 61 YR Male Hospitalization| R