DETERMINE HIV-1/2 AG/AB COMBO 25T
Report
- Report Number
- 1221359-2025-00403
- Event Type
- Malfunction
- Date Received
- October 1, 2025
- Date of Event
- March 20, 2025
- Report Date
- October 1, 2025
- Manufacturer
- ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
- Product Code
- MZF
- PMA / PMN Number
- BP120037
- Removal / Correction Number
- 1221359-07/31/2025-002-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT 0000939868 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 7D2648 / LOT 0000939868, DEVICE PART NUMBER 10732998 / LOT 944760. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE POSITIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT 0000939868 SHOWED THAT THE COMPLAINT RATE IS (B)(4). AS PART OF AN INVESTIGATION INTO PRELIMINARY FALSE REACTIVE RESULTS, ABBOTT FOUND THAT FOR KIT LOT 0000939868, THE ROOT CAUSE OF THE RESULTS WAS DUE TO A SUBCOMPONENT USED IN THE MANUFACTURE OF DETERMINE HIV-1/2 AG/AG COMBO. THE ISSUE HAS BEEN ISOLATED TO SPECIFIC LOTS OF THE SUBCOMPONENT, WHICH WERE USED TO MANUFACTURE SPECIFIC TEST KIT LOTS. THE NECESSARY ACTIONS TO PREVENT RECURRENCE HAVE BEEN TAKEN, AND THE CORRECTION OF PRODUCT IN THE FIELD WAS CONDUCTED.
THE CUSTOMER REPORTED FALSE POSITIVE RESULTS WITH THE DETERMINE HIV 1/2 AG/AB COMBO TEST PERFORMED ON VARIOUS DATES WITH A FINGERSTICK SAMPLE. THIS REPORT IS FOR PATIENT SIX (6) OF SIX (6). THE CUSTOMER REPORTED A FALSE POSITIVE RESULT WITH THE DETERMINE HIV 1/2 AG/AB COMBO TEST PERFORMED ON (B)(6) 2025. A FOURTH (4TH) GENERATION WITH REFLEX TO RNA QUALITATIVE TEST (BRAND UNKNOWN) WAS PERFORMED AND RETUNED A NEGATIVE RESULT. THERE IS NO PHYSICAL HARM TO THE PATIENT AS A RESULT OF THE RESULT. ALTHOUGH REQUESTED, NO INFORMATION REGARDING TREATMENT WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1658640 | DETERMINE HIV-1/2 AG/AB COMBO 25T | TEST, HIV DETECTION | MZF | ABBOTT DIAGNOSTICS SCARBOROUGH, INC. | 0000939868 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 28 YR | Female |