MCRYL+ VIO 27IN 3-0 S/A SH
Report
- Report Number
- 2210968-2025-11210
- Event Type
- Malfunction
- Date Received
- September 30, 2025
- Date of Event
- September 1, 2025
- Report Date
- December 4, 2025
- Manufacturer
- ETHICON INC.
- Product Code
- GAM
- UDI-DI
- 10705031046368
- PMA / PMN Number
- K050845
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
PRODUCT COMPLAINT # (B)(4). THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY ETHICON INC, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, ETHICON INC OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE. ADDITIONAL INFORMATION: H6, D9, H3. H3 INVESTIGATION SUMMARY: A FAILED SUTURE NEEDLE, LABELED AS (B)(4), WAS SUBMITTED TO HERRERA, STAFFORD AND ASSOCIATES, LLC (HSA) FOR FRACTOGRAPHIC EVALUATION ON SEPTEMBER 25, 2025. THE COMPONENT WAS IDENTIFIED AS PRODUCT CODE MCP316H. IT WAS REQUESTED THAT HSA ASSESS THE FRACTURE MODE OF THE FAILURE. ACCORDING TO THE INFORMATION, IT WAS REPORTED BREAKAGE NEEDLE. THE PRODUCT RECEIVED FOR ANALYSIS WAS MCP316H. VISUAL INSPECTION AND METALLURGICAL ANALYSIS WERE PERFORMED ON THE RETURNED PRODUCT. A FRACTURE WAS OBSERVED IN THE BODY OF THE NEEDLE. THE NEEDLE WAS RECEIVED IN TWO PIECES, ONLY ONE OF THE MATING FRACTURE SURFACES WAS EXAMINED FOR THIS EVALUATION. A SCANNING ELECTRON MICROSCOPE (SEM) WAS USED TO EXAMINE THE FRACTURE SURFACES AND SURROUNDING AREA OF THE NEEDLE. THE FRACTURE SURFACES WERE EXAMINED IN MULTIPLE LOCATIONS TO DETERMINE THE FRACTURE MODE. THE EVALUATION OF (B)(4) REVEALED THE FRACTURE WAS COMPOSED OF MICROVOID COALESCENCE, WHICH IS EVIDENCE OF A DUCTILE FRACTURE MODE. A CUP-AND-CONE SHAPED FRACTURE AND MACROSCOPIC PLASTIC DEFORMATION OBSERVED COINCIDENTAL TO THE FRACTURE PROVIDE ADDITIONAL EVIDENCE THAT THE FAILURE WAS INDUCED BY MECHANICAL DEFORMATION LEADING TO DUCTILE OVERLOAD. THIS WAS A DUCTILE FRACTURE. CORRECTED DATA: D4 UDI.
PRODUCT COMPLAINT #(B)(4). H6 COMPONENT CODE: G07002 - DEVICE NOT RETURNED THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803, PART 4 SUBPART B. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY ETHICON, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, ETHICON, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE. A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED DEVICE LOT, AND NO NON-CONFORMANCES WERE IDENTIFIED. TO DATE THE DEVICE HAS NOT BEEN RETURNED. IF THE DEVICE OR FURTHER DETAILS ARE RECEIVED AT A LATER DATE A SUPPLEMENTAL MEDWATCH WILL BE SENT. ADDITIONAL INFORMATION WAS REQUESTED, AND THE FOLLOWING WAS OBTAINED VIA: * PLEASE CONFIRM IF THE NEEDLE AND SUTURE OF THE SAME DEVICE BROKE DURING THIS PROCEDURE. --ONLY THE NEEDLE BROKE DURING THIS PROCEDURE * DID ANY NEEDLE PIECE FELL INSIDE OF THE PATIENT ? WAS IT RETRIEVED DURING THE SAME PROCEDURE. IF YES HOW WAS IT RETRIEVED? --NO * WERE THERE ANY PATIENT CONSEQUENCES? --NO
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, DOCTOR USED SUTURE FOR SUTURING. DURING USE, IT WAS FOUND THAT THE SUTURE AND NEEDLE WERE BROKEN. THE DOCTOR REPLACED THE SUTURE FOR SUTURING. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1672855 | MCRYL+ VIO 27IN 3-0 S/A SH | SUTURE, ABSORBABLE, SYNTHETIC | GAM | ETHICON INC. | 10791E | 10705031046368 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |