LUXSTAR UNIT MOUNT LIGHT
Report
- Report Number
- 3004517290-2011-00011
- Event Type
- Malfunction
- Date Received
- October 17, 2011
- Date of Event
- January 1, 2011
- Report Date
- September 26, 2011
- Manufacturer
- DENTAL EQUIPMENT LLC (DBA NEWBERG MANUFACTURING)
- Product Code
- EAZ
- Removal / Correction Number
- Z-2834-2011
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
THE LENS HEAT SHIELD/HOLDER FROM THE DENTAL LIGHT HAS A THREE PRONG LOCKING MECHANISM WHICH LOCKS THE LENS HEAT SHIELD/COVER INTO PLACE. THE LENS HEAT SHIELD/COVER HAS TO BE REMOVED BY THE END USER WHEN CLEANING THE LENS OR REPLACING THE HALOGEN BULB DURING ROUTINE OPERATOR MAINTENANCE. THE INSTALLATION INSTRUCTIONS, USE AND CARE INSTRUCTIONS, AND LABELING CLEARLY STATE TO ENSURE THE LENS HEAT SHIELD/HOLDER IS PROPERLY SECURE BY SNAPPING THE PART BACK INTO PLACE. DENTAL EQUIPMENT LLC INITIATED A RECALL ON (B)(6) 2011 TO ADD A TETHER TO THE LENS HEAT SHIELD/COVER TO KEEP IT FROM FALLING OFF THE DENTAL LIGHT IF IT IS NOT RE-INSTALLED PROPERLY. (B)(4).
A DOCTOR WAS PERFORMING ROUTINE MEDICAL TREATMENT TO A PT WHEN THE LENS HEAT SHIELD/HOLDER FROM HIS DENTAL LIGHT CAME OFF THE LIGHT AND FELL TO THE FLOOR. THERE WAS NO CONTACT WITH OR INJURY TO THE PT OR THE OPERATOR.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LUXSTAR UNIT MOUNT LIGHT | DENTAL OPERATING LIGHT | EAZ | DENTAL EQUIPMENT LLC (DBA NEWBERG MANUFACTURING) | UL1000 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |