FDA Adverse Event Malfunction Summary report: N

UNICEL® DXC SYNCHRON ® CHEMISTRY ANALYZER

MDR report key: 2317942 · Received November 1, 2011

Report

Report Number
2050012-2011-05072
Event Type
Malfunction
Date Received
November 1, 2011
Date of Event
June 29, 2011
Report Date
June 29, 2011
Manufacturer
BECKMAN COULTER INC.
Product Code
JJE
PMA / PMN Number
K042291
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Additional Manufacturer Narrative · 1

CHANGE FROM: THE CUSTOMER CONTACTED BECKMAN COULTER INC (BEC) TO REPORT ASFDAFD ASF ASDF GIVING "CUVETTE NOT DRY BEFORE FIRST REAGENT DISPENSE" ERRORS AND REAGENT PROBE B LEAK. TO: THE CUSTOMER CONTACTED BECKMAN COULTER INC(BEC) TO REPORT UNICEL DXC 800PRO SYNCHRON CHEMISTRY ANALYZER GAVE "CUVETTE NOT DRY BEFORE FIRST REAGENT DISPENSE" ERRORS AND REAGENT PROBE B LEAK.

Description of Event or Problem · 1

THE CUSTOMER CONTACTED BECKMAN COULTER INC(BEC) TO REPORT UNICEL DXC 800PRO SYNCHRON CHEMISTRY ANALYZER GAVE "CUVETTE NOT DRY BEFORE FIRST REAGENT DISPENSE" ERRORS AND REAGENT PROBE B LEAK.

Description of Event or Problem · 1

THE CUSTOMER CONTACTED BECKMAN COULTER INC(BEC) TO REPORT ASFDAFD ASF ASDF GIVING "CUVETTE NOT DRY BEFORE FIRST REAGENT DISPENSE" ERRORS AND REAGENT PROBE B LEAK. THE CUSTOMER WAS WEARING PPE WHEN SHE DISCOVERED THE LEAK. NO INJURY OR EXPOSURE TO THE FLUID WAS REPORTED AND PATIENT DIAGNOSIS OR TREATMENT WAS NOT IMPACTED BY THE EVENT. A BEC FIELD SERVICE ENGINEER (FSE) REPLACED THE A/B VALVE WHICH RESOLVED THE LEAKING ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNICEL® DXC SYNCHRON ® CHEMISTRY ANALYZER ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE JJE BECKMAN COULTER INC. DXC 800P N/A

Patients

Seq Age Sex Outcome Treatment
1