FDA Adverse Event Malfunction Summary report: N

MECTACER BIOLOX DELTA FEMORAL BALL HEAD

MDR report key: 23179116 · Received September 30, 2025

Report

Report Number
3005180920-2025-00918
Event Type
Malfunction
Date Received
September 30, 2025
Date of Event
September 2, 2025
Report Date
December 22, 2025
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
LZO
UDI-DI
07630030809262
PMA / PMN Number
K112115
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

BATCH REVIEW PERFORMED ON 22 DECEMBER 2025 BALL HEADS: MECTACER 01.29.215 BIOLOX DELTA CERAMIC BALL HEAD 12/14 Ø 40 SIZE XL +8 (K112115) LOT 2218785: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 02-NOV-2022. EXPIRATION DATE: 2027-10-13. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH NO SIMILAR REPORTED EVENTS DURING THE PERIOD OF REVIEW. LINER: MPACT 01.32.4052HCT FLAT PE HC LINER Ø40/G (K122641) LOT 2215506: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 06-SEP-2022. EXPIRATION DATE: 2027-08-21. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH NO SIMILAR REPORTED EVENTS DURING THE PERIOD OF REVIEW. CORRECTION: D1, D2, D4, G4, H4.

Additional Manufacturer Narrative · 0

BATCH REVIEW PERFORMED ON 15 SEPTEMBER 2025: LOT 2254377: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 18-AUG-2022. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, 4 SIMILAR EVENTS HAVE BEEN REPORTED ON THE SAME LOT DURING THE PERIOD OF REVIEW. VISUAL INSPECTION: THE PERFORMED ANALYSIS CONFIRMED THAT THE ANTIROTATIONAL INSERT OF THE TRIAL OFFSET HAS BEEN DISASSEMBLED FROM THE MAIN BODY DURING TRIAL REMOVAL FROM THE BONE. DURING EXTRACTION OF THE SYSTEM CONSISTING OF EXTENSION STEM, TRIAL OFFSET AND TRIAL KEEL ASSEMBLY FROM THE BONE, THE ELASTIC CLIPPING OF THE THREADED INSERT OF THE TRIAL OFFSET IS UNABLE TO WITHSTAND THE BLOWS AND DISENGAGES FROM THE OFFSET BODY ITSELF. A NEW MODIFICATION REQUEST HAS BEEN OPENED AND MANAGED TO IMPROVE THE LOCKING SYSTEM BETWEEN THE TWO COMPONENTS OF THE TRIAL OFFSET. ADDITIONAL IMPROVEMENT HAS BEEN MADE ON THE DESIGN AND PRODUCTION PROCESS IN ORDER TO OBTAIN THE ANTIROTATIONAL FEATURE MONOBLOCK WITH THE MAIN BODY INSTEAD OF TWO DIFFERENT COMPONENTS. CONCLUSION: THE ISSUE WAS LIKELY INFLUENCED BY A COMBINATION OF THE INHERENT MECHANICAL LIMITATIONS OF THE DESIGN VERSION INVOLVED AND THE FORCES APPLIED UNDER THE SPECIFIC CONDITIONS OF USE.

Description of Event or Problem · 0

THE ANTIROTATIONAL INSERT OF THE TRIAL OFFSET 5MM BROKE. BROKEN PART RETRIEVED AND SURGERY COMPLETED SUCCESSFULLY. APPROXIMATELY 45 MINUTES OF DELAY OCCURRED TO REMOVE THE BROKEN PART. THE SURGEON HAD TO BURR AROUND THE SITE TO GET ACCESS WITH A TRAUMA-THREADED REMOVAL TOOL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2014802 MECTACER BIOLOX DELTA FEMORAL BALL HEAD BIOLOX DELTA CERAMIC BALL HEAD 12/14 Ø 40 SIZE XL +8 LZO MEDACTA INTERNATIONAL SA 01.29.215 2218785 07630030809262

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Other