FDA Adverse Event Malfunction Summary report: N

VICRYL RAP UND 75CM M2 USP3/0 SGLE ARMED PS-2 PRIME

MDR report key: 23178865 · Received September 30, 2025

Report

Report Number
2210968-2025-11181
Event Type
Malfunction
Date Received
September 30, 2025
Date of Event
September 4, 2025
Report Date
December 5, 2025
Manufacturer
ETHICON INC.
Product Code
GAM
UDI-DI
10705031053014
PMA / PMN Number
K033746
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

PRODUCT COMPLAINT # (B)(4). THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY ETHICON INC, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, ETHICON INC OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE. ADDITIONAL INFORMATION: H6, D9, H3. H3 INVESTIGATION SUMMARY: A FAILED SUTURE NEEDLE, LABELED AS (B)(4), WAS SUBMITTED TO HERRERA, STAFFORD AND ASSOCIATES, LLC (HSA) FOR FRACTOGRAPHIC EVALUATION ON SEPTEMBER 25, 2025. THE COMPONENT WAS IDENTIFIED AS PRODUCT CODE 1960900. IT WAS REQUESTED THAT HSA ASSESS THE FRACTURE MODE OF THE FAILURE. ACCORDING TO THE INFORMATION, IT WAS REPORTED BREAKAGE NEEDLE. THE PRODUCT RECEIVED FOR ANALYSIS WAS W9923. VISUAL INSPECTION AND METALLURGICAL ANALYSIS WERE PERFORMED ON THE RETURNED PRODUCT. A FRACTURE WAS OBSERVED AT THE TIP OF THE NEEDLE. ONE SIDE OF THE NEEDLE WAS RECEIVED, THE MATING FRACTURE SURFACE WAS NOT PROVIDED FOR THIS EVALUATION. A SCANNING ELECTRON MICROSCOPE (SEM) WAS USED TO EXAMINE THE FRACTURE SURFACES AND SURROUNDING AREA OF THE NEEDLE. THE FRACTURE SURFACES WERE EXAMINED IN MULTIPLE LOCATIONS TO DETERMINE THE FRACTURE MODE. THE EVALUATION OF (B)(4) REVEALED THE FRACTURE WAS COMPOSED OF MICROVOID COALESCENCE, WHICH IS EVIDENCE OF A DUCTILE FRACTURE MODE. MECHANICAL DAMAGE OBSERVED COINCIDENTAL TO THE FRACTURE PROVIDES ADDITIONAL EVIDENCE THAT THE FAILURE WAS INDUCED BY MECHANICAL DEFORMATION LEADING TO DUCTILE OVERLOAD. THIS WAS A DUCTILE FRACTURE.

Additional Manufacturer Narrative · 0

PRODUCT COMPLAINT #: (B)(4). THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY ETHICON INC, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, ETHICON INC OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE. ADDITIONAL INFORMATION: H6, D9, H3. H3 INVESTIGATION SUMMARY: THE PRODUCT WAS RETURNED TO ETHICON FOR EVALUATION. ONE NEEDLE SUTURE COMBINATION AND ONE NEEDLE SUTURE PIECE WERE RECEIVED FOR ANALYSIS. PRODUCT CODE W9923. DURING THE VISUAL INSPECTION OF THE RETURNED SAMPLES, NO ISSUES RELATED TO NEEDLE BREAKAGE WERE OBSERVED. HOWEVER, IT WAS NOTED THAT THE CURVATURE OF ONE NEEDLE WAS UNUSUAL DUE TO THIS ONE WAS NOTED TO BE DISTORTED PROBABLY CAUSED BY A SURGICAL INSTRUMENT. IN ADDITION, CRIMPING MARKS AND BODY FLUIDS WERE FOUND ON THE NEEDLES. AS PART OF THE ETHICON QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. ALTHOUGH NO CONCLUSION COULD BE REACHED ON THE CAUSE OF THE REPORTED EVENT. TO AVOID THIS KIND OF DAMAGE: GRASP THE NEEDLE IN AN AREA ONE-THIRD (1/3) TO ONE-HALF (1/2) OF THE DISTANCE FROM THE ATTACHMENT END TO THE POINT. THE EVENT DESCRIBED COULD NOT BE CONFIRMED AS NO BREAKAGE NEEDLE WAS NOTED. ALTHOUGH NO PRODUCT DEFECT WAS IDENTIFIED, THERE MAY HAVE BEEN OTHER CIRCUMSTANCES OR ISSUES THAT OCCURRED DURING THE USE OF THE DEVICE THAT COULD NOT BE REPLICATED DURING THE LABORATORY ANALYSIS. A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED DEVICE LOT NUMBER AND NO NON-CONFORMANCES / MANUFACTURING IRREGULARITIES WERE IDENTIFIED.

Additional Manufacturer Narrative · 0

PRODUCT COMPLAINT # (B)(4) DATE SENT TO THE FDA: 9/30/2025. H6 COMPONENT CODE: G07002 - PENDING EVALUATION OF RETURNED DEVICE. THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY ETHICON, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, ETHICON, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE. A DEVICE HAS BEEN RECEIVED; HOWEVER IT HAS NOT YET BEEN EVALUATED. ANY FURTHER INFORMATION DERIVED FROM THE EVALUATION WILL BE SUBMITTED IN A SUPPLEMENTAL 3500A FORM. D4/G4: DEVICE IS NOT DISTRIBUTED IN THE UNITED STATES BUT IS SIMILAR TO DEVICE MARKETED IN THE USA. THEREFORE, (01) GTIN IS NOT AVAILABLE.

Additional Manufacturer Narrative · 0

PRODUCT COMPLAINT # (B)(4). THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY ETHICON INC, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, ETHICON INC OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE. ADDITIONAL INFORMATION: H6, D9, H3. H3 INVESTIGATION SUMMARY: A FAILED SUTURE NEEDLE, LABELED AS (B)(4), WAS SUBMITTED TO HERRERA, STAFFORD AND ASSOCIATES, LLC (HSA) FOR FRACTOGRAPHIC EVALUATION ON SEPTEMBER 25, 2025. THE COMPONENT WAS IDENTIFIED AS PRODUCT CODE 1960900. IT WAS REQUESTED THAT HSA ASSESS THE FRACTURE MODE OF THE FAILURE. ACCORDING TO THE INFORMATION, IT WAS REPORTED BREAKAGE NEEDLE. THE PRODUCT RECEIVED FOR ANALYSIS WAS W9923. VISUAL INSPECTION AND METALLURGICAL ANALYSIS WERE PERFORMED ON THE RETURNED PRODUCT. A FRACTURE WAS OBSERVED AT THE TIP OF THE NEEDLE. ONE SIDE OF THE NEEDLE WAS RECEIVED, THE MATING FRACTURE SURFACE WAS NOT PROVIDED FOR THIS EVALUATION. A SCANNING ELECTRON MICROSCOPE (SEM) WAS USED TO EXAMINE THE FRACTURE SURFACES AND SURROUNDING AREA OF THE NEEDLE. THE FRACTURE SURFACES WERE EXAMINED IN MULTIPLE LOCATIONS TO DETERMINE THE FRACTURE MODE. THE EVALUATION OF (B)(4) REVEALED THE FRACTURE WAS COMPOSED OF MICROVOID COALESCENCE, WHICH IS EVIDENCE OF A DUCTILE FRACTURE MODE. MECHANICAL DAMAGE OBSERVED COINCIDENTAL TO THE FRACTURE PROVIDES ADDITIONAL EVIDENCE THAT THE FAILURE WAS INDUCED BY MECHANICAL DEFORMATION LEADING TO DUCTILE OVERLOAD. THIS WAS A DUCTILE FRACTURE. CORRECTED DATA: D4 UDI.

Description of Event or Problem · 0

IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2025 AND SUTURE WAS USED. IT WAS REPORTED THAT DURING SURGERY, THE NEEDLE TIP HAD BROKEN AS ATTACHED DURING SEWING. CHANGED ANOTHER ONE TO COMPLETE SURGERY. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2014793 VICRYL RAP UND 75CM M2 USP3/0 SGLE ARMED PS-2 PRIME SUTURE, ABSORBABLE, SYNTHETIC GAM ETHICON INC. AU7013 10705031053014

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown