ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2011-08121
- Event Type
- Injury
- Date Received
- November 1, 2011
- Report Date
- October 11, 2011
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER
Narratives
THE PUMP HAS BEEN RETURNED TO ANIMAS. EVALUATION HAS NOT YET BEEN COMPLETED. WHEN EVALUATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSION CAN BE MADE AT THIS TIME. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
FOLLOW-UP #1 11/29/2011-DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(6) 2011 WITH THE FOLLOWING FINDINGS:AN EZPRIME OPERATION WAS PERFORMED WITH NO ISSUES. THE PUMP PASSED THE FORCE SENSOR CALIBRATION TEST. A REVIEW OF THE TOTAL DAILY DOSE HISTORY SHOWED THAT THE DAILY INSULIN DELIVERY TOTALS CORRECTLY REFLECTED THE USER PROGRAMMED BASAL RATES. THE BLACK BOX SHOWED MULTIPLE LOSS OF PRIME WARNINGS, ONE OF WHICH WAS UNACKNOWLEDGED FOR TWO HOURS. A 29 HOUR FLOW ACCURACY TEST WAS PERFORMED WITHOUT INTERRUPTIONS AND THE PUMP WAS FOUND TO BE DELIVERING WITHIN SPECIFICATIONS. THE PUMP WAS OPEND AND THE FORCE SENSOR ASSEMBLY WAS FOUND TO BE DAMAGED; 2531779-03/24/2010-003-R.
IT WAS REPORTED THAT THE PATIENT'S BLOOD GLUCOSE LEVEL ROSE TO 506MG/DL WITHOUT SYMPTOMS OF HYPERGLYCEMIA. THE REPORTER INDICATED THAT THE PUMP EMITTED 3 LOSS OF PRIME WARNINGS EARLIER IN THE MORNING, BUT WAS UNABLE TO DETERMINE HOW LONG THE PATIENT WAS WITHOUT INSULIN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ONETOUCH PING INSULIN PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 12 YR | Life Threatening |