FDA Adverse Event Malfunction Summary report: N

I-STAT TROPONIN (CTNI) CARTRIDGE

MDR report key: 2317850 · Received October 14, 2011

Report

Report Number
2245578-2011-00265
Event Type
Malfunction
Date Received
October 14, 2011
Date of Event
July 11, 2011
Report Date
September 23, 2011
Manufacturer
ABBOTT POINT OF CARE INC.
Product Code
MMI
Removal / Correction Number
APOC2011-007
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DETAILS WILL BE PROVIDED WITH THE ACTION THAT WILL BE CONDUCTED IN COOPERATION WITH THE U.S. FOOD AND DRUG ADMINISTRATION.

Description of Event or Problem · 1

ON (B)(6) 2011, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT TROPONIN CARTRIDGES THAT YIELDED DISCREPANT RESULT ON A PT IN THE ED. I-STAT: DATE: (B)(6) 2011, 16:16, RESULT: 0.08 NG/ML. DATE: (B)(6) 2011, 16:18, RESULT: 0.08 NG/ML, DATE: (B)(6) 2011, 16:34, RESULT: 0.10 NG/ML. (B)(6) (LAB): DATE: (B)(6) 2011, 15:48, RESULT: <0.01 NG/ML. THE SUSPECTED DISCREPANT RESULTS WERE NOT RELEASED TO A PHYSICIAN OR CAREGIVER. BASED ON THE INFO AVAILABLE AT THE TIME, THERE WERE NO INJURIES ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 I-STAT TROPONIN (CTNI) CARTRIDGE CTNI CARTRIDGE MMI ABBOTT POINT OF CARE INC. NA T11096

Patients

Seq Age Sex Outcome Treatment
1 UNK