FDA Adverse Event
Injury
Summary report: N
INFINION? CX
MDR report key: 23177934
·
Received September 30, 2025
Report
- Report Number
- 3006630150-2025-08142
- Event Type
- Injury
- Date Received
- September 30, 2025
- Date of Event
- July 16, 2025
- Report Date
- September 30, 2025
- Manufacturer
- BOSTON SCIENTIFIC NEUROMODULATION CORPORATION
- Product Code
- LGW
- PMA / PMN Number
- P030017
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- ID, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
Additional Manufacturer Narrative · 0
BLOCK B3: APPROXIMATED BASED ON THE DATE THE MANUFACTURER BECAME AWARE OF THE EVENT. ADDITIONAL SUSPECT MEDICAL DEVICE COMPONENT INVOLVED IN THE EVENT: PRODUCT FAMILY: SCS-LINEAR LEADS. UPN: M365SC2317500. MODEL: SC-2317-50. SERIAL: (B)(6). BATCH: 7072589. UDI: (B)(4).
Description of Event or Problem · 0
IT WAS REPORTED THAT THE PATIENTS SPINAL CORD STIMULATOR (SCS) LEADS HAD HIGH IMPEDANCE CAUSING INADEQUATE STIMULATION. THE PATIENT UNDERWENT A LEAD REPLACEMENT PROCEDURE AND WAS DOING WELL POSTOPERATIVELY. THE EXPLANTED LEADS WERE DISCARDED BY THE MEDICAL FACILITY AND WILL NOT BE RETURNED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1772970 | INFINION? CX | STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF | LGW | BOSTON SCIENTIFIC NEUROMODULATION CORPORATION | SC-2317-50 | 5093547 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 54 YR | Female | Required Intervention |