ALINITY I STAT HIGH SENSITIVE TROPONIN-I REAGENT KIT
Report
- Report Number
- 3005094123-2025-00484
- Event Type
- Malfunction
- Date Received
- September 29, 2025
- Date of Event
- September 5, 2025
- Report Date
- September 29, 2025
- Manufacturer
- ABBOTT IRELAND DIAGNOSTICS DIVISION
- Product Code
- MMI
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
ALL AVAILABLE PATIENT INFORMATION WAS INCLUDED. ADDITIONAL PATIENT DETAILS ARE NOT AVAILABLE. THIS REPORT IS BEING FILED ON AN INTERNATIONAL PRODUCT, LIST NUMBER 08P13, THAT HAS A SIMILAR PRODUCT DISTRIBUTED IN THE US, LIST NUMBER 04Z21 (ALINITY I STAT HIGH SENSITIVITY TROPONIN-I), AND A 510K NUMBER OF K202525. THE COMPLAINT INVESTIGATION FOR FALSELY ELEVATED ALINITY I STAT HIGH SENSITIVE TROPONIN-I RESULTS INCLUDED A REVIEW OF COMPLAINT TEXT, A SEARCH FOR SIMILAR COMPLAINTS, TRENDING DATA, LABELING, AND DEVICE HISTORY RECORDS. THE OVERALL PERFORMANCE OF ALINITY I STAT HIGH SENSITIVE TROPONIN-I REAGENTS IN THE FIELD WAS REVIEWED USING DATA GATHERED FROM CUSTOMERS WORLDWIDE. THE PATIENT MEDIAN VALUES FOR THE LOT ARE WITHIN THE ESTABLISHED LIMITS AND COMPARABLE TO THE HISTORICAL REAGENT LOT PERFORMANCE. A REVIEW OF TRACKING AND TRENDING FOR THE PRODUCT AND THE LIKELY CAUSE LOT DID NOT IDENTIFY AN INCREASE IN COMPLAINT ACTIVITY. DEVICE HISTORY RECORD REVIEW DID NOT SHOW ANY POTENTIAL NON-CONFORMANCES OR DEVIATIONS. LABELING WAS REVIEWED AND FOUND TO ADEQUATELY ADDRESS THE ISSUE. BASED ON THE INFORMATION PROVIDED AND ABBOTT DIAGNOSTICS¿ COMPLAINT INVESTIGATION, NO SYSTEMIC ISSUE OR DEFICIENCY FOR THE ALINITY I STAT HIGH SENSITIVE TROPONIN-I, LOT NUMBER 77406UD00, WAS IDENTIFIED.
THE CUSTOMER OBSERVED A FALSELY ELEVATED ALINITY I STAT HIGH SENSITIVE TROPONIN-I RESULT FOR A 36-YEAR-OLD FEMALE PATIENT. THE FOLLOWING DATA WAS PROVIDED (CUSTOMER¿S REFERENCE RANGE IS <16 NG/L): SAMPLE ID (B)(6) 9 INITIAL RESULT WAS 73.6, REPEAT RESULT WAS <8, REPEAT RESULT, ON A DIFFERENT ANALYZER, WAS <8 NG/L. THE PATIENT WAS RECOLLECTED 2 HOURS LATER, AND THE RESULT WAS <8 NG/L. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1796985 | ALINITY I STAT HIGH SENSITIVE TROPONIN-I REAGENT KIT | IMMUNOASSAY METHOD, TROPONIN SUBUNIT | MMI | ABBOTT IRELAND DIAGNOSTICS DIVISION | 77406UD00 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 36 YR | Female | ALNTY I PROCESSING MODU, 03R65-01, (B)(6). |