FDA Adverse Event Injury Summary report: N

MINIMED MIO ADVANCE

MDR report key: 23164343 · Received September 29, 2025

Report

Report Number
8021545-2025-02546
Event Type
Injury
Date Received
September 29, 2025
Date of Event
August 23, 2025
Report Date
November 18, 2025
Manufacturer
UNOMEDICAL A/S
Product Code
FPA
PMA / PMN Number
K173879
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

E1: (B)(6).

Additional Manufacturer Narrative · 0

THE INITIAL MDR WITH MANUFACTURING REPORT NUMBER (8021545-2025-02546), WAS SUBMITTED ON 29-SEP-2025. UPON COMPLETION OF THE INVESTIGATION, THE MANUFACTURING DATE WAS UPDATED AS 04-NOV-2024. ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY. COMPLAINT INVESTIGATION RESULTS: A COMPLAINT INVESTIGATION HAS BEEN INITIATED UNDER COMPLAINT INVESTIGATION CHILD RECORD (B)(4). THE BATCH 6010006 IN QUESTION WAS MANUFACTURED AT THE OSTED SITE. THRESHOLD ANALYSIS: A QUERY WAS RUN ON 18-NOV-2025 AGAINST "FINAL REPORTING DECISION EQUAL "SERIOUS INJURY" AND "DEATH", "LOT NUMBER" CRITERIA EQUAL "6010006". THE COUNT OF COMPLAINT IS 1 WHICH IS BELOW 3 SO NO FURTHER STATISTICAL TRENDING ANALYSIS IS REQUIRED. DEVICE HISTORY RECORD (DHR) REVIEW: THE LOT 6010006 WAS MANUFACTURED ACCORDING TO WORK INSTRUCTION (WI) [80] APPENDIX 1 BATCHCARD FOR PRODUCTION OF PACKAGING ROOM ON 04-NOV-2024, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DHR SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WAS IDENTIFIED, NOR MAINTENANCE EVENTS WERE RECORDED RELATED TO COMPLAINT CODE. TEST RESULTS: NO SAMPLE WAS PROVIDED. CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: AS A RESULT OF THE FOLLOWING: NO NON-CONFORMANCE (NC) RAISED DURING PRODUCTION RELATED TO COMPLAINT CODE, NO TREND IDENTIFIED, NO FURTHER ACTIONS ARE REQUIRED. THIS COMPLAINT WILL NOT REQUIRE FURTHER ROOT CAUSE INVESTIGATION NOR CORRECTIVE AND PREVENTIVE ACTION (CAPA) PLAN. THEREFORE, THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKET SURVEILLANCE ACTIVITIES.

Description of Event or Problem · 0

REFERENCE NUMBER (B)(4). EVENT OCCURRED IN CANADA. IT WAS REPORTED THAT THE PATIENT FACED AN EVENT OF HIGH BLOOD GLUCSOSE LEVEL ON (B)(6) 2025 ALSO, THE PATIENT WAS SICK, TIRED, HAD PAIN AND CRAMPS. THEREFORE, THE PATIENT WENT TO EMERGENCY ROOM FOR ONE DAY WITH BLOOD GLUCOSE LEVEL OF 34.08 MMOL/L. DURING HOSPITALIZATION, THE PATIENT WAS TREATED WITH PUMP AND INTRAVENOUSLY. NO FURTHER INFORMATION AVAILABLE.

Description of Event or Problem · 0

TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1839483 MINIMED MIO ADVANCE UNO MIO ADV. GREY 60/6 SC1 10-PK CAN FPA UNOMEDICAL A/S MMT-242A600 6010006

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Hospitalization| R