DAVINCI XI
Report
- Report Number
- 2955842-2025-39805
- Event Type
- Malfunction
- Date Received
- September 27, 2025
- Date of Event
- September 4, 2025
- Report Date
- September 4, 2025
- Manufacturer
- INTUITIVE SURGICAL, INC
- Product Code
- NAY
- PMA / PMN Number
- K131861
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IA, US
- Reporter Occupation
- NURSE
- Health Professional
- Yes
Narratives
AN INVESTIGATION WAS COMPLETED TO DETERMINE THE CAUSE OF THIS REPORTED EVENT. AN INTUITIVE SURGICAL, INC. (ISI) FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER SITE TO FURTHER INVESTIGATE THE REPORTED EVENT. SITE REPORTED THAT THE PREVIOUSLY REPORTED ISSUE HAS RECURRED. DURING THE PROCEDURE, THE SYSTEM EXPERIENCED RECOVERABLE FAULTS THAT ESCALATED TO NON-RECOVERABLE ERRORS. THE SYSTEM WAS RESTARTED, AND ALL FIBER CONNECTIONS WERE CHECKED. LOG REVIEW INDICATED AN ERROR RELATED TO THE PERSONALITY MODULE AUDIO / VIDEO (PMAV), AND A 307-ERROR REPORTED SUGGESTED A POTENTIAL PMAV ISSUE. PART HISTORY COULD NOT BE VERIFIED DUE TO MISSING PMAV SERIAL NUMBER. ALL PMAV CONNECTIONS WERE CONFIRMED SECURE, AND THE SURGEON PLANNED TO CONTINUE WITH SUBSEQUENT CASES. FSE VISITED THE SITE AND OBSERVED THE ERRORS. FSE REPLACED THE PMAV. THE SYSTEM WAS TESTED AND VERIFIED AS READY FOR USE. ISI DID RECEIVE A DA VINCI PRODUCT INVOLVED WITH THIS COMPLAINT TO PERFORM FAILURE ANALYSIS. THE PMAV WAS ANALYZED AND FOUND THAT THE ISSUE WAS CONFIRMED BUT NOT REPLICATED BY THE FAILURE ANALYSIS. VISUAL INSPECTION REVEALED NO ISSUES RELATED TO THE REPORTED FAILURE. SYSTEM LOGS REPORTED ERROR 307 AND 23 NODE PMAV INDICATING A HARDWARE FAULT IN WHEEL COMMUNICATION, CONFIRMING A FIELD OCCURRENCE. THE PMAV WAS TESTED ON A GOLDEN PRINTED CIRCUIT ASSEMBLY SYSTEM, FUNCTIONING CORRECTLY WITH NO ERRORS, GOOD IMAGE QUALITY, AND PASSING ALL I/O TESTS. AFTER 30 POWER CYCLES AND EXTENDED OPERATION, THE UNIT PERFORMED NORMALLY WITHOUT ANY ISSUES. THE PROBABLE ROOT CAUSE IN THIS CASE CANNOT BE DETERMINED AS THE FAILURE ANALYSIS DID NOT REPLICATE THE CUSTOMER REPORTED ISSUE. THIS ISSUE WAS RESOLVED BY REPLACING THE PMAV.
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED HYSTERECTOMY SURGICAL PROCEDURE, THE CUSTOMER INFORMED INTUITIVE SURGICAL, INC. (ISI) TECHNICAL SUPPORT ENGINEER (TSE) THAT REPEATED ERRORS 23 AND 307 OCCURRED. THE CUSTOMER RESTARTED THE SYSTEM AND CONFIRMED THAT ALL CONNECTIONS TO THE PERSONALITY MODULE AUDIO / VIDEO (PMAV) WERE TIGHT AND PROPERLY SEATED. THE TSE CONFIRMED ERROR 23 IN THE LOGS POINTING TO THE PMAV. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1663715 | DAVINCI XI | VISION SIDE SYSTEM | NAY | INTUITIVE SURGICAL, INC | 381121-10 | N/A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Female | DA VINCI INSTRUMENTS AND ACCESSORIES |