FDA Adverse Event Malfunction Summary report: N

KOS - RAPID MICROWAVE HISTOPROCESSOR

MDR report key: 2315430 · Received October 11, 2011

Report

Report Number
1000391282-2011-00002
Event Type
Malfunction
Date Received
October 11, 2011
Date of Event
September 7, 2011
Report Date
October 6, 2011
Manufacturer
MILESTONE S.R.L.
Product Code
KPA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Description of Event or Problem · 1

DURING THE STANDARD USE OF THE UNIT, WHILE IN THE PARAFFIN-WAX STEP, PARAFFIN HEATED TO A TEMPERATURE THAT MELTED THE FORTY-FIVE CASSETTES. AS A CONSEQUENCE, THE BIOPSY TISSUE INSIDE THE CASSETTES WERE DAMAGED AND A SATISFACTORY DIAGNOSIS WAS NOT POSSIBLE. THE UNIT WAS OPERATING ON A POWER SUPPLY SOURCE AND GROUND CONDITION NOT ADEQUATE BASED ON THE PROVIDED POWER REQUIREMENTS DESIGNATED ON THE PRE-INSTALLATION GUIDE AND USER MANUAL PROVIDED TO THE USER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 KOS - RAPID MICROWAVE HISTOPROCESSOR PROCESSOR, TISSUE, AUTOMATED KPA MILESTONE S.R.L. KOS

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention