K5-2500 CASTROVIEJO FORCEPS 0.12MM
Report
- Report Number
- 3010692967-2025-00076
- Event Type
- Malfunction
- Date Received
- September 25, 2025
- Report Date
- September 25, 2025
- Manufacturer
- CORZA OPTHALMOLOGY (KATENA PRODUCTS INC.)
- Product Code
- HTD
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TH
- Reporter Occupation
- 501
Narratives
ALTHOUGH IT REMAINS UNCLEAR WHETHER THE DISCOLORATION IN THE SUBMITTED IMAGE IS RUST, THIS REPORT IS BEING FILED DUE TO CONCERNS THAT THE UNIDENTIFIED PHENOMENON MAY INCREASE THE RISK OF INFECTION. THE REPORTER NOTED THAT NO CLEARER IMAGE OF THE AFFECTED INSTRUMENT IS AVAILABLE AND INSTEAD PROVIDED A PHOTO OF A SEPARATE INSTRUMENT EXHIBITING A SIMILAR ISSUE. THE IMAGE SHOWS DISCOLORATION RATHER THAN RUST AT THE TIP. CORZA HAS PROVIDED THE CUSTOMER WITH THE OFFICIAL CARE AND HANDLING GUIDE, WHICH OUTLINES THE DIFFERENCES BETWEEN RUST AND DISCOLORATION ON STAINLESS STEEL INSTRUMENTS AND OFFERS GUIDANCE ON PREVENTION. THE REPORTER CONFIRMED THAT THE SUSPECTED INSTRUMENT WILL NOT BE RETURNED FOR EVALUATION, AND NO LOT NUMBER WAS PROVIDED. ADDITIONALLY, THE USER CLAIMED THAT THE DISCOLORATION WAS OBSERVED BEFORE THE INSTRUMENT CAME INTO CONTACT WITH THE PATIENT, AND THE INSTRUMENT WAS NEVER USED. A REVIEW OF PAST COMPLAINTS INDICATES THAT RUST TYPICALLY RESULTS FROM IMPROPER CLEANING AND STORAGE PRACTICES. DISCOLORATION MAY OCCUR WHEN UNAPPROVED DETERGENTS ARE USED DURING THE CLEANING PROCESS. THE INSTRUMENT'S VALIDATED INSTRUCTIONS FOR USE (IFU LT-2) INCLUDE DETAILED GUIDANCE ON APPROVED CLEANING AGENTS AND PROCEDURES. THE IFU ALSO STATES THAT THE INSTRUMENT SHOULD BE THOROUGHLY INSPECTED PRIOR TO USE.
ON AUGUST 26, 2025, A DISTRIBUTOR FROM THAILAND CONTACTED CORZA CUSTOMER SERVICE VIA EMAIL TO INQUIRE ABOUT POTENTIAL CAUSES OF RUST ON STAINLESS STEEL INSTRUMENTS. THE EMAIL INCLUDED AN IMAGE SHOWING A K5-2500 CASTROVIEJO FORCEPS IN USE DURING A PROCEDURE. THE PHOTO WAS CAPTURED FROM A VIDEO MONITOR DISPLAYING THE LIVE FEED FROM THE DOCTOR'S SCOPE. ALTHOUGH THE IMAGE IS BLURRY, SLIGHT DISCOLORATION IS VISIBLE ON THE INSTRUMENT'S JAW, WHICH MAY RESEMBLE RUST. CORZA REQUESTED A CLEARER IMAGE FROM THE CUSTOMER AND ASKED THAT THE INSTRUMENT BE RETURNED FOR EVALUATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 323410 | K5-2500 CASTROVIEJO FORCEPS 0.12MM | CASTROVIEJO FORCEPS | HTD | CORZA OPTHALMOLOGY (KATENA PRODUCTS INC.) | K5-2500 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |