I-STAT TROPONIN (CTNI) CARTRIDGE
Report
- Report Number
- 2245578-2011-00258
- Event Type
- Malfunction
- Date Received
- October 6, 2011
- Date of Event
- August 4, 2011
- Report Date
- September 23, 2011
- Manufacturer
- ABBOTT POINT OF CARE INC.
- Product Code
- MMI
- Removal / Correction Number
- APOC2011-007
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WV, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
(B)(4). MFR REPORT #2245578-2011-00217 THROUGH 2245578-2011-00222, 2245578-2011-00228 AND 2245578-2011-00243 THROUGH 2245578-2011-00259 ARE FROM A SINGLE USER SITE. (B)(4). DETAILS WILL BE PROVIDED WITH THE ACTION THAT WILL BE CONDUCTED IN COOPERATION WITH THE U.S. FOOD AND DRUG ADMINISTRATION.
ON (B)(6) 2011, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT TROPONIN CARTRIDGES THAT YIELDED DISCREPANT RESULT ON A PATIENT. I-STAT TESTING RESULTS (NG/ML): (B)(6) 2011, 17:12, COLLECTION; (B)(6) 2011, 17:28, TESTING WITH RESULT OF 0.02; (B)(6) 2011, 18:43, 90 MINUTE COLLECTION; (B)(6) 2011, 19:05, 90 MINUTE TESTING WITH RESULT OF 0.07. NO REPEAT TESTING PERFORMED. CENTAUR LAB CTNI RESULT <0.006. THE SUSPECT DISCREPANT RESULTS WERE RELEASED TO A PHYSICIAN OR CAREGIVER. BASED ON THE INFORMATION AVAILABLE AT THE TIME, THERE WERE NO INJURIES ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | I-STAT TROPONIN (CTNI) CARTRIDGE | CTNI CARTRIDGE | MMI | ABBOTT POINT OF CARE INC. | NA | T11096 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 19 YR |