FDA Adverse Event Injury Summary report: N

ASPIRE PLATFORM

MDR report key: 2314535 · Received October 21, 2011

Report

Report Number
1223483-2011-00001
Event Type
Injury
Date Received
October 21, 2011
Date of Event
January 1, 2010
Report Date
October 21, 2011
Manufacturer
PALOMAR MEDICAL TECHNOLOGIES, INC.
Product Code
GEX
PMA / PMN Number
080567
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
SW
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

AT THE TIME OF DEVICE INSTALLATION THE DISTRIBUTOR CONFIRMED BOTH DEVICES WERE WORKING WITHIN SPECIFICATIONS. ADDITIONALLY, (B)(6) POST INSTALLATION THE DISTRIBUTOR ALSO VERIFIED BOTH DEVICES WERE OPERATING WITHIN SPECIFICATIONS. BOTH DEVICE FACILITIES AS WELL AS THE TREATING PHYSICIAN BELIEVE THE DEVICES OPERATED AS INTENDED OVER THE COURSE OF THE YEAR. AFTER REVIEWING SEVERAL OF THE CASES WITH PHYSICIANS TRAINED IN PLASTIC SURGERY, IT WAS BELIEVED THESE INCIDENTS OCCURRED AS A RESULT OF POOR SUCTION TECHNIQUE AND WERE UNRELATED TO THE LASER SYSTEM. THE PHYSICIAN IS NO LONGER USING THESE DEVICES. ACCORDING TO THE USER FACILITIES NONE OF THE PATIENTS WERE HOSPITALIZED AND DID NOT EXPERIENCE INFECTION. PATIENT'S SKIN DISCOLORATION IS IMPROVING POST PEELS AND MICRODERM ABRASION. SCARRING IS ALSO IMPROVING WITH SILICONE PAD APPLICATIONS.

Description of Event or Problem · 1

OVER THE COURSE OF A YEAR MULTIPLE PATIENTS WERE TREATED BY THE SAME PHYSICIAN IN (B)(6). COMPLICATIONS WERE REPORTED BETWEEN TWO DIFFERENT LOCATIONS EACH HAVING THEIR OWN DEVICE. REPORTED COMPLICATIONS INCLUDED: POOR RESULTS AND CONTOUR DEFORMITIES IN APPROXIMATELY 35 PATIENTS. NECROSIS AND SCARRING IN APPROXIMATELY 10-15 PATIENTS. SKIN COLOR CHANGES IN APPROXIMATELY 20-30 PATIENTS AND ONE PATIENT EXPERIENCED BURN. BOTH DEVICES OPERATED WITHIN SPECIFICATION AND PERFORMED AS INTENDED. WITH THE EXCEPTION OF THE BURN, POOR SUCTION TECHNIQUE POST LASER TREATMENT IS THOUGHT TO BE THE CAUSE OF COMPLICATIONS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ASPIRE PLATFORM GEX PALOMAR MEDICAL TECHNOLOGIES, INC. 0300-1000

Patients

Seq Age Sex Outcome Treatment
1 Other TUMESCENT/ANESTHESIA