LIBERTY
Report
- Report Number
- 2428983-2011-00003
- Event Type
- Malfunction
- Date Received
- October 14, 2011
- Date of Event
- March 23, 2011
- Report Date
- October 14, 2011
- Manufacturer
- GF HEALTH PRODUCTS, INC./ DBA BASIC AMERICAN
- Product Code
- FNL
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
CONCLUSION FROM (B)(4): "IT HAS NOT BEEN POSSIBLE TO RE-CREATE THE FAILURE. EVALUATING THE RETURNED UNITS AND THE DESIGN IT IS EVIDENT A DOUBLE FAULT HAS OCCURRED. THE X-RAY IS SHOWING A DEFECTIVE MOTOR CABLE WITH A BAD FEEDBACK (END STOP) WIRE. THE MOST PLAUSIBLE EXPLANATION IS MISSING THE END STOP SIGNAL FROM THE ACTUATOR DUE TO THE MOTOR CABLE AND INPUT OF CONSTANT ACTIVATION FROM HANDSET OR OTHER EXTERNAL DEVICE (LAYING ON HAND PENDANT, UNINTENDED ACTIVATION OF BUTTON IN HOLSTER OR BETWEEN MATTRESS/FRAME/WALL). (B)(4) IS CURRENTLY SELLING APPROX (B)(4) PCS OF THESE SYSTEMS ANNUALLY, WITH SIMILAR CONFIGURATION ((B)(4)) AND HAVE HAD NO SUCH FAILURE PREVIOUSLY REPORTED. (B)(4). DESPITE THE VERY UNPLEASANT ODOR WHICH MAY BE ASSOCIATED WITH FIRE, THERE WAS NO SUCH THING AND THE PRODUCT FAILED IN A SAFE MODE AS INTENDED." (B)(4). ADD'L MODEL#: 0277020-650 MOTOR CABLE, (B)(4).
ON OR ABOUT (B)(6) 2011, (B)(6), ALLEGED THAT THE ACTUATOR CASING FROM GF HEALTH PRODUCTS, INC. (GFHP) LIBERTY MODEL BED SMELLED LIKE IT WAS ON FIRE. THE ACTUATOR CASING AND MOTOR ASSEMBLY IS MANUFACTURED BY (B)(4). THIS COMPONENT IS USED FOR THE LIBERTY MODEL BED MANUFACTURED BY GFHP. ON (B)(4) 2011, GFHP ISSUED AN RGA ((B)(4)) AND DROP SHIPPED THE COMPONENTS DIRECTLY FROM THE CUSTOMER ((B)(6)) TO THE MFR ((B)(4).) FOR TESTING. IT SHOULD ALSO BE NOTED, THERE WAS NO REPORT INJURY ASSOCIATED WITH THIS ACTUATOR MALFUNCTION. A WORKER DETECTED A "HOT" SMELL, AND THE BED WAS TAKEN OUT OF USE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIBERTY | LIBERTY BED | FNL | GF HEALTH PRODUCTS, INC./ DBA BASIC AMERICAN | LA27-U094-00 ACTUATOR | 07/27/2010 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |