Report Number
9617032-2025-01690
Event Type
Malfunction
Date Received
September 25, 2025
Date of Event
August 25, 2025
Report Date
October 21, 2025
Manufacturer
BECTON, DICKINSON AND COMPANY (BD)
Product Code
JKA
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FI
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY BD HAD NOT RECEIVED ANY SAMPLES OR PHOTOS FOR INVESTIGATION. A TOTAL OF 100 RETAINED SAMPLES FROM EACH LOT NUMBER PROVIDED WERE VISUALLY INSPECTED, AND NO GEL DEFECT RELATED TO CLOGGED INSTRUMENTS WAS FOUND. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOTS, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THIS COMPLAINT HAS NOT BEEN CONFIRMED FOR THE LOT 5071480, FOR THE INDICATED FAILURE MODE: CLOGGED INSTRUMENTATION. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR THE IDENTIFICATION OF EMERGING TRENDS.

Description of Event or Problem · 0

REPORT 3 OF 3: IT WAS REPORTED THAT WHILE USING BD VACUTAINER® SST¿ II ADVANCE PLUS BLOOD COLLECTION TUBES, AN UNSPECIFIED NUMBER OF TUBES CAUSED CLOGGED SAMPLE PROBES ON THE ATELLICA ANALYZERS. NO PATIENT IMPACT REPORTED.

Description of Event or Problem · 0

REPORT 3 OF 3. IT WAS REPORTED THAT WHILE USING BD VACUTAINER® SST¿ II ADVANCE PLUS BLOOD COLLECTION TUBES, AN UNSPECIFIED NUMBER OF TUBES CAUSED CLOGGED SAMPLE PROBES ON THE ATELLICA ANALYZERS. NO PATIENT IMPACT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2430064 BD VACUTAINER® SST¿ II ADVANCE PLUS BLOOD COLLECTION TUBES BLOOD SPECIMEN COLLECTION DEVICE JKA BECTON, DICKINSON AND COMPANY (BD) 5071480

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown