FDA Adverse Event
Malfunction
Summary report: N
I-STAT TROPONIN (CTNI) CARTRIDGE
MDR report key: 2313420
·
Received October 3, 2011
Report
- Report Number
- 2245578-2011-00219
- Event Type
- Malfunction
- Date Received
- October 3, 2011
- Date of Event
- September 4, 2011
- Report Date
- September 23, 2011
- Manufacturer
- ABBOTT POINT OF CARE INC.
- Product Code
- MMI
- Removal / Correction Number
- APOC2011-007
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WV, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). DETAILS WILL BE PROVIDED WITH THE ACTION THAT WILL BE CONDUCTED IN COOPERATION WITH THE U.S. FOOD AND DRUG ADMINISTRATION.
Description of Event or Problem · 1
ON (B)(6) 2011, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT TROPONIN CARTRIDGES THAT YIELDED DISCREPANT RESULT OF 0.14 NG/ML ON A PT. I-STAT RESULTS (NG/ML): ON (B)(6) 2011 11:09 INITIAL COLLECTION. ON (B)(6) 2011 11:16 INITIAL TESTING WITH RESULT OF 0.14. ON (B)(6) 2011 12:38 90 MINUTE COLLECTION. ON (B)(6) 2011 12:41 90 MINUTE TESTING WITH RESULT OF <0.01. CENTAUR LAB CTNI RESULT: 0.012. THE SUSPECTED DISCREPANT RESULTS WERE RELEASED TO A PHYSICIAN OR CAREGIVER. BASED ON THE INFORMATION AVAILABLE AT THE TIME, THERE WERE NO INJURIES ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | I-STAT TROPONIN (CTNI) CARTRIDGE | CTNI CARTRIDGE | MMI | ABBOTT POINT OF CARE INC. | NA | T11140 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 71 YR |