I-STAT TROPONIN (CTNI) CARTRIDGE
Report
- Report Number
- 2245578-2011-00228
- Event Type
- Malfunction
- Date Received
- October 3, 2011
- Date of Event
- September 1, 2011
- Report Date
- September 23, 2011
- Manufacturer
- ABBOTT POINT OF CARE INC.
- Product Code
- MMI
- Removal / Correction Number
- APOC2011-007
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WV, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). DETAILS WILL BE PROVIDED WITH THE ACTION THAT WILL BE CONDUCTED IN COOPERATION WITH THE U.S. FOOD AND DRUG ADMINISTRATION.
ON (B)(6) 2011, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT TROPONIN CARTRIDGES THAT YIELDED DISCREPANT RESULT OF 0.10 NG/ML ON A PT. I-STAT RESULTS (NG/ML): ON (B)(6) 2011 21:00 INITIAL COLLECTION. ON (B)(6) 2011 21:08 INITIAL TESTING WITH RESULT OF 0.10. ON (B)(6) 2011 22:30 90 MINUTE COLLECTION. ON (B)(6) 2011 22:37 90 MINUTE TESTING WITH RESULT OF 0.02. NO REPEAT TESTING PERFORMED ON THE I-STAT. CENTAUR LAB CTNI RESULT: <0.006. THE SUSPECTED DISCREPANT RESULTS WERE RELEASED TO A PHYSICIAN OR CAREGIVER. BASED ON THE INFORMATION AVAILABLE AT THE TIME, THERE WERE NO INJURIES ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | I-STAT TROPONIN (CTNI) CARTRIDGE | CTNI CARTRIDGE | MMI | ABBOTT POINT OF CARE INC. | NA | T11140 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 23 YR |