FDA Adverse Event Malfunction Summary report: N

I-STAT TROPONIN (CTNI) CARTRIDGE

MDR report key: 2313399 · Received October 3, 2011

Report

Report Number
2245578-2011-00217
Event Type
Malfunction
Date Received
October 3, 2011
Date of Event
September 4, 2011
Report Date
September 23, 2011
Manufacturer
ABBOTT POINT OF CARE, INC.
Product Code
MMI
Removal / Correction Number
APOC2011-007
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WV, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DETAILS WILL BE PROVIDED WITH THE ACTION THAT WILL BE CONDUCTED IN COOPERATION WITH THE U.S. FOOD AND DRUG ADMINISTRATION.

Description of Event or Problem · 1

ON (B)(6) 2011, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT TROPONIN CARTRIDGES THAT YIELDED DISCREPANT RESULT OF 0.12 NG/ML ON A PT. I-STAT RESULTS (NG/ML): ON (B)(6) 2011 11:57 INITIAL COLLECTION. ON (B)(6) 2011 11:56 INITIAL TESTING WITH RESULT OF <0.01. ON (B)(6) 2011 13:34 90 MINUTE COLLECTION. ON (B)(6) 2011 13:34 90 TESTING WITH RESULT OF 0.12. NO REPEAT TESTING PERFORMED ON THE I-STAT. LAB RESULT: <0.006 (CENTAUR). THE SUSPECTED DISCREPANT RESULTS WERE RELEASED TO A PHYSICIAN OR CAREGIVER. BASED ON THE INFORMATION AVAILABLE AT THE TIME, THERE WERE NO INJURIES ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 I-STAT TROPONIN (CTNI) CARTRIDGE CTNI CARTRIDGE MMI ABBOTT POINT OF CARE, INC. NA T11140

Patients

Seq Age Sex Outcome Treatment
1 39 YR