FDA Adverse Event Injury Summary report: N

QUICK SET

MDR report key: 23130135 · Received September 24, 2025

Report

Report Number
3003442380-2025-14062
Event Type
Injury
Date Received
September 24, 2025
Date of Event
August 22, 2025
Report Date
August 27, 2025
Manufacturer
UNOMEDICAL DEVICES S.A. DE C.V.
Product Code
FPA
PMA / PMN Number
K991759
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

E1:PATIENT CITY: (B)(6). PATIENT COUNTRY: UNITED ARAB EMIRATES. ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY: COMPLAINT INVESTIGATION RESULTS: A COMPLAINT INVESTIGATION HAS BEEN INITIATED UNDER COMPLAINT INVESTIGATION CHILD RECORD (B)(4). THE BATCH 6003383, IN QUESTION WAS MANUFACTURED AT THE REYNOSA SITE. TEST RESULTS: VISUAL TEST ACCORDING TO WORK INSTRUCTION (WI) VERSION 3.0 ON REFERENCE SAMPLES, 10 SAMPLES OUT 10 SAMPLES PASSED THE TEST. FUNCTIONAL FLOW TEST 1 ACCORDING TO WI VERSION 2.0 ON REFERENCE SAMPLES, 10 SAMPLES OUT 10 SAMPLES PASSED THE TEST. FUNCTIONAL LEAK TEST 2 ACCORDING TO WI VERSION 2.0 ON REFERENCE SAMPLES, 10 SAMPLES OUT 10 SAMPLES PASSED THE TEST. SEE ATTACHMENT DATABASE (B)(4) COMPLAINT TEST REPORT.PDF FOR THE DETAILS. DEVICE HISTORY RECORD (DHR) REVIEW: THE LOT 6003383 WAS MANUFACTURED ACCORDING TO THE WI VERSION 77 MANUFACTURED IN MULTIVAC M12 ON 18-SEP-2023, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DEVICE HISTORY RECORD (DHR) SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WAS IDENTIFIED, NOR MAINTENANCE EVENTS WERE RECORDED RELATED TO COMPLAINT CODE. CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: AS A RESULT OF THE FOLLOWING: NO PHYSICAL SAMPLES WERE RETURNED, NO NON-CONFORMANCE (NC) RAISED DURING PRODUCTION RELATED TO COMPLAINT CODE, NO TREND IDENTIFIED FOR THE LOT IN QUESTION, NO FURTHER ACTIONS ARE REQUIRED. THIS COMPLAINT WILL NOT REQUIRE FURTHER ROOT CAUSE INVESTIGATION NOR CORRECTIVE AND PREVENTIVE ACTION (CAPA) PLAN. THEREFORE, THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKET SURVEILLANCE ACTIVITIES.

Description of Event or Problem · 0

REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED ARAB EMIRATES. IT WAS REPORTED THAT THE PATIENT WAS HOSPITALIZED ON (B)(6) 2025 DUE TO BENT CANNULA LEADING TO HYPERGLYCEMIA. THE BLOOD GLUCOSE LEVEL WAS 380 MG/DL AT THE TIME OF EVENT AND THE PATIENT GOT TREATED WITH INTRAVENOUS INSULIN. THE LENGTH OF HOSPITALIZATION WAS LESS THAN 24 HOURS. THE PATIENT ALSO HAD SYMPTOMS LIKE VOMITING AND ABDOMINAL PAIN. NO FURTHER INFORMATION AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
542194 QUICK SET UNO QUICK-SET 60/6 SC1 MECA FPA UNOMEDICAL DEVICES S.A. DE C.V. MMT-399A 6003383

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Required Intervention