GIRAFFE OMNIBED
Report
- Report Number
- 1121732-2011-00004
- Event Type
- Death
- Date Received
- October 21, 2011
- Date of Event
- September 22, 2011
- Report Date
- October 21, 2011
- Manufacturer
- OHMEDA MEDICAL
- Product Code
- FMZ
- PMA / PMN Number
- K101788
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- LA, US
- Reporter Occupation
- NURSE
Narratives
THE HOSPITAL IS CURRENTLY CHANGING THEIR CLEANING POLICIES AND IMPLEMENTING CHANGES OF BOTH HOW AND WHEN TO CLEAN THE MICROENVIRONMENTS. THERE WAS NO ALLEGATION OF A DEVICE MALFUNCTION. THE HOSPITAL INITIALLY CONTACTED GE HEALTHCARE TO INQUIRE ABOUT THE CLEANING PROCEDURES FOR THE GIRAFFE OMNIBED. BASED ON THE INFO AVAILABLE AT THIS TIME, IT IS UNK WHETHER THE EVENT IS ATTRIBUTABLE TO THE DEVICE. GE HEALTHCARE'S INVESTIGATION IS ONGOING. A FOLLOW UP REPORT WILL BE SUBMITTED ONCE THE INVESTIGATION HAS BEEN COMPLETED. THE INFANT AGE WAS (B)(6). THE SERIAL NUMBER OF THE DEVICE INVOLVED IN THIS CASE COULD NOT BE OBTAINED FROM THE FACILITY; CONSEQUENTLY, NEITHER THE SERIAL NUMBER, NOR THE DEVICE MANUFACTURE DATE CAN BE DETERMINED. THE 510(K) NUMBER PROVIDED IS THE MOST CURRENT CLEARED PREMARKET NOTIFICATION FOR THE GIRAFFE OMNIBED.
THE HOSPITAL REPORTED THAT AN INFANT BEING TREATED IN A GIRAFFE OMNIBED HAD DIED WITH A POSITIVE CULTURE FOR (B)(6). THE INFANT WAS REPORTEDLY BEING TREATED WITH GREATER THAN 80% HUMIDITY IN THE MICROENVIRONMENT AND BED TEMPERATURES BETWEEN 37 AND 38 DEGREES CELSIUS. THE HOSPITAL PERFORMED RANDOM CULTURES ON A GIRAFFE OMNIBED AND AN OSCILLATOR WITH NEGATIVE RESULTS. THE HOSPITAL IS REPORTEDLY UNDERGOING EXTENSIVE RENOVATION, INCLUDING THE NICU (NEONATAL INTENSIVE CARE UNIT), AND A COMMENT WAS MADE AS TO WHETHER THE CONSTRUCTION PROJECT MAY HAVE AFFECTED THE WATER. THE TWO INFANTS WERE REPORTEDLY LOCATED IN AREAS OF THE NICU THAT SHARED A COMMON SINK. THE HOSPITAL STAFF IS BEING RE-EDUCATED IN HAND WASHING PROTOCOL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | GIRAFFE OMNIBED | NEONATAL INCUBATOR | FMZ | OHMEDA MEDICAL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Death |